Author:
Handa Takashi,Negoro Makoto,Miyachi Shigeru,Sugita Kenichiro
Abstract
✓ The pressure in 47 arteries feeding 21 arteriovenous malformations (AVM's) was investigated during transarterial embolization using a Tracker-18 microcatheter. On average, systolic pressure increased by 22 mm Hg. In AVM's with single or few feeders, embolization was usually achieved well; in contrast, giant AVM's with multiple feeders and a large arteriovenous shunt were poorly embolized. However, large AVM's with well-demarcated components may be reduced by embolization to an appropriate size for surgery or stereotactic radiation therapy. It was found that the feeding artery pressure increased significantly more in well-embolized than in poorly embolized cases. Measurement of the feeding artery pressure clarified the hemodynamics of AVM's and facilitated more successful embolization.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Cited by
42 articles.
订阅此论文施引文献