Inflammation and obstruction of distal catheter slits in ventriculoperitoneal shunts: likely role of graphite

Author:

Del Bigio Marc R.123,Sidhu Ravinder K.4,Kazina Colin J.5,Serletis Demitre56

Affiliation:

1. Department of Pathology, University of Manitoba;

2. Children’s Hospital Research Institute of Manitoba;

3. SharedHealth Manitoba;

4. Manitoba Institute for Materials, University of Manitoba;

5. Section of Neurosurgery, University of Manitoba and Health Sciences Centre; and

6. Manitoba Neurosurgery Laboratory, Children’s Hospital Research Institute of Manitoba, Winnipeg, Mannitoba, Canada

Abstract

OBJECTIVETissue reactions that contribute to obstruction of peritoneal catheters in ventriculoperitoneal shunt systems are not well characterized. Several recent rapid obstructions in children prompted a retrospective quality assurance review.METHODSThe authors conducted a detailed investigation of 22 surgically explanted peritoneal shunt catheters and 8 autopsy cases with documented distal shunt obstruction. Patients’ medical histories were reviewed, and the catheters and/or tissues were subjected to conventional histological and immunohistochemical evaluations. In addition, 3 cases were subjected to electron microscopic examination including elemental analysis.RESULTSThe majority of symptomatic obstructions were associated with distal slit catheters (17 slit, 3 open-end, and 2 unknown type). Among the autopsy cases, deaths were attributed to shunt failure in 2 cases of slit catheter obstruction, 1 case of open-end catheter obstruction, and 1 case of catheter withdrawal from the peritoneal cavity. The early tissue response consisted of a predominantly T lymphocyte accumulation with phagocytosis of graphite particles by macrophages. This is associated with proliferation of fibroblasts, mesothelial cells, and blood vessels, which can grow through the slits and occlude the catheter lumen. As the inflammation subsides after approximately 1 year, the tissue plug becomes collagenized and calcified.CONCLUSIONSThis study, supported by experimental literature in other organ systems, indicates that graphite used to coat the slit openings of distal catheters from ventriculoperitoneal shunts likely predisposes to obstruction. Neurosurgeons and manufacturers should consider the potential negative consequences of this shunt design. The authors concur with previous recommendations that slit-valve distal catheters should not be used for ventriculoperitoneal shunting unless they can be proven safe and efficacious in a controlled trial.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Genetics,Animal Science and Zoology

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3