A novel miniature robotic guidance device for stereotactic neurosurgical interventions: preliminary experience with the iSYS1 robot

Author:

Minchev Georgi1,Kronreif Gernot2,Martínez-Moreno Mauricio1,Dorfer Christian1,Micko Alexander1,Mert Aygül1,Kiesel Barbara1,Widhalm Georg1,Knosp Engelbert1,Wolfsberger Stefan1

Affiliation:

1. Department of Neurosurgery, Medical University of Vienna, Vienna; and

2. Austrian Center of Medical Innovation and Technology, Wiener Neustadt, Austria

Abstract

OBJECTIVE Robotic devices have recently been introduced in stereotactic neurosurgery in order to overcome the limitations of frame-based and frameless techniques in terms of accuracy and safety. The aim of this study is to evaluate the feasibility and accuracy of the novel, miniature, iSYS1 robotic guidance device in stereotactic neurosurgery. METHODS A preclinical phantom trial was conducted to compare the accuracy and duration of needle positioning between the robotic and manual technique in 162 cadaver biopsies. Second, 25 consecutive cases of tumor biopsies and intracranial catheter placements were performed with robotic guidance to evaluate the feasibility, accuracy, and duration of system setup and application in a clinical setting. RESULTS The preclinical phantom trial revealed a mean target error of 0.6 mm (range 0.1–0.9 mm) for robotic guidance versus 1.2 mm (range 0.1–2.6 mm) for manual positioning of the biopsy needle (p < 0.001). The mean duration was 2.6 minutes (range 1.3–5.5 minutes) with robotic guidance versus 3.7 minutes (range 2.0–10.5 minutes) with manual positioning (p < 0.001). Clinical application of the iSYS1 robotic guidance device was feasible in all but 1 case. The median real target error was 1.3 mm (range 0.2–2.6 mm) at entry and 0.9 mm (range 0.0–3.1 mm) at the target point. The median setup and instrument positioning times were 11.8 minutes (range 4.2–26.7 minutes) and 4.9 minutes (range 3.1–14.0 minutes), respectively. CONCLUSIONS According to the preclinical data, application of the iSYS1 robot can significantly improve accuracy and reduce instrument positioning time. During clinical application, the robot proved its high accuracy, short setup time, and short instrument positioning time, as well as demonstrating a short learning curve.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Genetics,Animal Science and Zoology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3