Author:
Ohe Makoto,Moridaira Hiroshi,Inami Satoshi,Takeuchi Daisaku,Nohara Yutaka,Taneichi Hiroshi
Abstract
OBJECTIVEInstrumentation failure caused by the loosening of pedicle screws (PSs) in patients with osteoporosis is a serious problem after spinal surgery. The addition of a thin hydroxyapatite (HA) surface coating applied by using a sputtering process was reported recently to be a promising method for providing bone conduction around an implant without a significant risk of coating-layer breakage. In this study, the authors evaluated the biomechanical and histological features of the bone-implant interface (BII) of PSs with a thin HA coating in an in vivo porcine osteoporotic spine model.METHODSThree types of PSs (untreated/standard [STPS], sandblasted [BLPS], and HA-coated [HAPS] PSs) were implanted into the thoracic and lumbar spine (T9–L6) of 8 mature Clawn miniature pigs (6 ovariectomized [osteoporosis group] and 2 sham-operated [control group] pigs). The spines were harvested from the osteoporosis group at 0, 2, 4, 8, 12, or 24 weeks after PS placement and from the control group at 0 or 24 weeks. Their bone mineral density (BMD) was measured by peripheral quantitative CT. Histological evaluation of the BIIs was conducted by performing bone volume/tissue volume and bone surface/implant surface measurements. The strength of the BII was evaluated with extraction torque testing.RESULTSThe BMD decreased significantly in the osteoporosis group (p < 0.01). HAPSs exhibited the greatest mean extraction peak torque at 8 weeks, and HAPSs and BLPSs exhibited significantly greater mean torque than the STPSs at 12 weeks (p < 0.05). The bone surface/implant surface ratio was significantly higher for HAPSs than for STPSs after 2 weeks (p < 0.05), and bonding between bone and the implant surface was maintained until 24 weeks with no detachment of the coating layer. In contrast, the bone volume/tissue volume ratio was significantly higher for HAPSs than for BLPSs or STPSs only at 4 weeks.CONCLUSIONSUsing PSs with a thin HA coating applied using a sputtering process strengthens bonding at the BII, which might improve early implant fixation after spinal surgery for osteoporosis. However, the absence of increased bone mass around the screw remains a concern; prescribing osteoporosis treatment to improve bone quality might be necessary to prevent fractures around the screws.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Reference46 articles.
1. In vitro biomechanical comparison of pedicle screws, sublaminar hooks, and sublaminar cables;Hitchon;J Neurosurg,2003
2. Pedicular fixation in the osteoporotic spine: a pilot in vivo study on long-term ovariectomized sheep;Aldini;J Orthop Res,2002
3. A comparative study on screw loosening in osteoporotic lumbar spine fusion between expandable and conventional pedicle screws;Wu;Arch Orthop Trauma Surg,2012
4. Histometric evaluation of bone around titanium implants with different surface treatments in rats exposed to cigarette smoke inhalation;Correa;Clin Oral Implants Res,2009
5. Bone-hydroxyapatite interface in retrieved hydroxyapatite-coated titanium implants: a clinical and histologic report;Piattelli;Int J Oral Maxillofac Implants,1999
Cited by
29 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献