Abstract
Background
Wearable devices are valuable assessment tools for patient outcomes in contexts such as clinical trials. To be successfully deployed, however, participants must be willing to wear them. Another concern is that usability studies are rarely published, often fail to test devices beyond 24 hours, and need to be repeated frequently to ensure that contemporary devices are assessed.
Objective
This study aimed to compare multiple wearable sensors in a real-world context to establish their usability within an older adult (>50 years) population.
Methods
Eight older adults wore seven devices for a minimum of 1 week each: Actigraph GT9x, Actibelt, Actiwatch, Biovotion, Hexoskin, Mc10 Biostamp_RC, and Wavelet. Usability was established through mixed methods using semistructured interviews and three questionnaires, namely, the Intrinsic Motivation Inventory (IMI), the System Usability Scale (SUS), and an acceptability questionnaire. Quantitative data were reported descriptively and qualitative data were analyzed using deductive content analysis. Data were then integrated using triangulation.
Results
Results demonstrated that no device was considered optimal as all scored below average in the SUS (median, IQR; min-max=57.5, 12.5; 47.5-63.8). Hexoskin was the lowest scored device based on the IMI (3.6; 3.4-4.5), while Biovotion, Actibelt, and Mc10 Biostamp_RC achieved the highest median results on the acceptability questionnaire (3.6 on a 6-point Likert scale). Qualitatively, participants were willing to accept less comfort, less device discretion, and high charging burdens if the devices were perceived as useful, namely through the provision of feedback for the user. Participants agreed that the purpose of use is a key enabler for long-term compliance. These views were particularly noted by those not currently wearing an activity-tracking device. Participants believed that wrist-worn sensors were the most versatile and easy to use, and therefore, the most suitable for long-term use. In particular, Actiwatch and Wavelet stood out for their comfort. The convergence of quantitative and qualitative data was demonstrated in the study.
Conclusions
Based on the results, the following context-specific recommendations can be made: (1) researchers should consider their device selection in relation to both individual and environmental factors, and not simply the primary outcome of the research study; (2) if researchers do not wish their participants to have access to feedback from the devices, then a simple, wrist-worn device that acts as a watch is preferable; (3) if feedback is allowed, then it should be made available to help participants remain engaged; this is likely to apply only to people without cognitive impairments; (4) battery life of 1 week should be considered as a necessary feature to enhance data capture; (5) researchers should consider providing additional information about the purpose of devices to participants to support their continued use.
Reference53 articles.
1. Using e-technologies in clinical trials
2. Health at hand: A systematic review of smart watch uses for health and wellness
3. BatesMIEEE Pulse Magazine201603142019-03-20New Trends in Clinical Trialshttps://pulse.embs.org/march-2016/new-trends-in-clinical-trials/
4. Twenty-four Hours of Sleep, Sedentary Behavior, and Physical Activity with Nine Wearable Devices
5. International Organization for Standardization2018-03-20ISO 13407:1999(en) Human-Centred Design Processes for Interactive Systemshttps://www.iso.org/obp/ui/#iso:std:iso:13407:ed-1:v1:en
Cited by
67 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献