Validation of Heart Rate Extracted From Wrist-Based Photoplethysmography in the Perioperative Setting: Prospective Observational Study

Author:

Mestrom EvelineORCID,Deneer RubenORCID,Bonomi Alberto GORCID,Margarito JennyORCID,Gelissen JosORCID,Haakma ReinderORCID,Korsten Hendrikus H MORCID,Scharnhorst VolkherORCID,Bouwman R ArthurORCID

Abstract

Background Measurement of heart rate (HR) through an unobtrusive, wrist-worn optical HR monitor (OHRM) could enable earlier recognition of patient deterioration in low acuity settings and enable timely intervention. Objective The goal of this study was to assess the agreement between the HR extracted from the OHRM and the gold standard 5-lead electrocardiogram (ECG) connected to a patient monitor during surgery and in the recovery period. Methods In patients undergoing surgery requiring anesthesia, the HR reported by the patient monitor’s ECG module was recorded and stored simultaneously with the photopletysmography (PPG) from the OHRM attached to the patient’s wrist. The agreement between the HR reported by the patient’s monitor and the HR extracted from the OHRM’s PPG signal was assessed using Bland-Altman analysis during the surgical and recovery phase. Results A total of 271.8 hours of data in 99 patients was recorded simultaneously by the OHRM and patient monitor. The median coverage was 86% (IQR 65%-95%) and did not differ significantly between surgery and recovery (Wilcoxon paired difference test P=.17). Agreement analysis showed the limits of agreement (LoA) of the difference between the OHRM and the ECG HR were within the range of 5 beats per minute (bpm). The mean bias was –0.14 bpm (LoA between –3.08 bpm and 2.79 bpm) and –0.19% (LoA between –5 bpm to 5 bpm) for the PPG- measured HR compared to the ECG-measured HR during surgery; during recovery, it was –0.11 bpm (LoA between –2.79 bpm and 2.59 bpm) and –0.15% (LoA between –3.92% and 3.64%). Conclusions This study shows that an OHRM equipped with a PPG sensor can measure HR within the ECG reference standard of –5 bpm to 5 bpm or –10% to 10% in the perioperative setting when the PPG signal is of sufficient quality. This implies that an OHRM can be considered clinically acceptable for HR monitoring in low acuity hospitalized patients.

Publisher

JMIR Publications Inc.

Subject

Cardiology and Cardiovascular Medicine,Health Informatics,Computer Science Applications

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3