Recent Trends in Performance and Current State of Creatinine Assays

Author:

Killeen Anthony A.1,Ashwood Edward R.1,Ventura Christina B.1,Styer Patricia1

Affiliation:

1. From the Department of Laboratory Medicine and Pathology, University of Minnesota, Minneapolis (Dr Killeen); ARUP Laboratories, Inc, University of Utah, Salt Lake City (Dr Ashwood); and the College of American Pathologists, Northfield, Illinois (Ms Ventura and Dr Styer).

Abstract

Context.—Measured plasma or serum creatinine concentration is a primary component of equations used to calculate estimated glomerular filtration rate (eGFR). In recent years, most assay manufacturers have adopted creatinine calibration procedures that are traceable to the National Institute of Standards and Technology's Standard Reference Material 967. Objectives.—To examine the current performance of creatinine assays, to compare changes in assay performance since 2003, and to examine the reliability of laboratory eGFR calculations. Design.—Serum samples spiked with different concentrations of creatinine were analyzed by participating laboratories in the College of American Pathologists' LN24 survey. Participants' reported values were compared against values measured by liquid chromatography-isotope dilution mass spectrometry. Participants were asked to calculate the eGFR for certain samples, and results were compared with those obtained from the 4-parameter Modification of Diet in Renal Disease equation. Results.—Biases among current creatinine methods are in the range of −5% to 10%, compared with −7% to 34% seen in a 2003 study. This degree of bias in eGFR calculations is of clinical significance only for concentrations near the cut points used to stage chronic kidney disease. Approximately 20% of laboratories report eGFR values that exceed ±1 mL/min per 1.73 m2 from the expected eGFR using the 4-parameter Modification of Diet in Renal Disease equation. Conclusions.—Since 2003, there have been improvements in the performance of creatinine assays, which appear to be related to the adoption of standard reference materials for calibration. The effect of the observed method biases in clinical practice now appears minimal. Laboratories should continue to monitor the accuracy of eGFR calculations.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3