Approaches for Performance Verification Toward Standardization of Peripheral Blood Regulatory T-Cell Detection by Flow Cytometry

Author:

Liu Mei12,Liu Jin-Peng12,Wang Pan12,Fu Ya-Jing12,Zhao Min1,Jiang Yong-Jun12,Zhang Zi-Ning12,Shang Hong12

Affiliation:

1. From NHC Key Laboratory of AIDS Immunology - China Medical University (M Liu, J-P Liu, Wang, Fu, Jiang, Zhang, Shang), National Clinical Research Center for Laboratory Medicine (M Liu, J-P Liu, Wang, Fu, Zhao, Jiang, Zhang, Shang), and the Department of Laboratory Medicine (Zhao, Shang), The First Hospital of China Medical University, Shenyang, China

2. Units of Medical Laboratory (M Liu, J-P Liu, Wang, Fu, Jiang, Zhang, Shang) and Key Laboratory of AIDS Immunology (J-P Liu, Wang, Fu, Jiang, Zhang, Shang), Chinese Academy of Medical Sciences, Shenyang, China

Abstract

Context.— Regulatory T-cell (Treg) detection in peripheral blood, based on flow cytometry, is invaluable for diagnosis and treatment of immune-mediated diseases. However, there is a lack of reliable methods to verify the performance, which is pivotal towards standardization of the Tregs assay. Objective.— To conduct standardization studies and verify the performance of 3 commercially available reagent sets for the Tregs assay based on flow cytometry and agreement analysis for Treg detection across the different reagent sets. Design.— The analytical performance of Tregs assay using reagent sets supplied by 3 manufacturers was evaluated after establishing the gating strategy and determining the optimal antibody concentration. Postcollection sample stability was evaluated, as well as the repeatability, reproducibility, reportable range, linearity, and assay carryover. Agreement between the different assays was assessed via Bland-Altman plots and linear regression analysis. The relationship between the frequency of CD4+CD25+CD127low/− Tregs and CD4+CD25+Foxp3+ Tregs was evaluated. Results.— The postcollection sample stability was set at 72 hours after collection at room temperature. The accuracy, repeatability, reproducibility, and accuracy all met the requirements for clinical analysis. Excellent linearity, with R2 ≥0.9 and no assay carryover, was observed. For reportable range, a minimum of 1000 events in the CD3+CD4+ gate was required for Tregs assay. Moreover, the results for Tregs labeled by antibodies from the 3 manufacturers were in good agreement. The percentage of CD4+CD25+CD127low/− Tregs was closely correlated with CD4+CD25+Foxp3+ Tregs. Conclusions.— This is the first study to evaluate systematically the measurement performance of Tregs in peripheral blood by flow cytometry, which provides a practical solution to verifying the performance of flow cytometry–based immune monitoring projects in clinical practice.

Publisher

Archives of Pathology and Laboratory Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3