Creatinine Measurement: State of the Art in Accuracy and Interlaboratory Harmonization

Author:

Greg Miller W.1,Myers Gary L.1,Ashwood Edward R.1,Killeen Anthony A.1,Wang Edward1,Thienpont Linda M.1,Siekmann Lothar1

Affiliation:

1. From the Virginia Commonwealth University, Richmond (Dr Miller); Centers for Disease Control and Prevention, Atlanta, Ga (Dr Myers); ARUP Laboratories, University of Utah, Salt Lake City (Dr Ashwood); University of Minnesota, Minneapolis (Dr Killeen); College of American Pathologists, Northfield, Ill (Dr Wang); Ghent University, Gent, Belgium (Dr Thienpont); and the Institute of Clinical Biochemi

Abstract

Abstract Context.—The National Kidney Disease Education Program recommends calculating glomerular filtration rate from serum creatinine concentration. Accurate creatinine measurements are necessary for this calculation. Objective.—To evaluate the state of the art in measuring serum creatinine, as well as the ability of a proficiency testing program to measure bias for individual laboratories and method peer groups. Design.—A fresh-frozen, off-the-clot pooled serum specimen plus 4 conventional specimens were sent to participants in the College of American Pathologists Chemistry Survey for assay of creatinine. Creatinine concentrations were assigned by isotope dilution mass spectrometry reference measurement procedures. Participants.—Clinical laboratories with an acceptable result for all 5 survey specimens (n = 5624). Results.—The fresh frozen serum (FFS) specimen had a creatinine concentration of 0.902 mg/dL (79.7 μmol/L). Mean bias for 50 instrument-method peer groups varied from −0.06 to 0.31 mg/dL (−5.3 to 27.4 μmol/L), with 30 (60%) of 50 peer groups having significant bias (P < .001). The bias variability was related to instrument manufacturer (P ≤ .001) rather than method type (P = .02) with 24 (63%) of 38 alkaline picric acid methods and with 6 (50%) of 12 enzymatic methods having significant biases. Two conventional specimens had creatinine concentrations of 0.795 and 2.205 mg/dL (70.3 and 194.9 μmol/L) and had apparent survey biases significantly different (P < .001) from that of the FFS specimen for 34 (68%) and 35 (70%) of 50 peer groups, respectively. Conclusions.—Thirty of 50 peer groups had significant bias for creatinine. Bias was primarily associated with instrument manufacturer, not with type of method used. Proficiency testing using a commutable specimen measured participant bias versus a reference measurement procedure and provided trueness surveillance of instrument-method peer groups.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3