Author:
van der Jagt M. Annerie,Briaire Jeroen J.,Verbist Berit M.,Frijns Johan H.M.
Abstract
The HiFocus Mid-Scala (MS) electrode array has recently been introduced onto the market. This precurved design with a targeted mid-scalar intracochlear position pursues an atraumatic insertion and optimal distance for neural stimulation. In this study we prospectively examined the angular insertion depth achieved and speech perception outcomes resulting from the HiFocus MS electrode array for 6 months after implantation, and retrospectively compared these with the HiFocus 1J lateral wall electrode array. The mean angular insertion depth within the MS population (n = 96) was found at 470°. This was 50° shallower but more consistent than the 1J electrode array (n = 110). Audiological evaluation within a subgroup, including only postlingual, unilaterally implanted, adult cochlear implant recipients who were matched on preoperative speech perception scores and the duration of deafness (MS = 32, 1J = 32), showed no difference in speech perception outcomes between the MS and 1J groups. Furthermore, speech perception outcome was not affected by the angular insertion depth or frequency mismatch.
Subject
Speech and Hearing,Sensory Systems,Otorhinolaryngology,Physiology
Cited by
29 articles.
订阅此论文施引文献