Continuous Subcutaneous Glucose Monitoring in Children With Type 1 Diabetes

Author:

Chase H. Peter1,Kim Laura M.1,Owen Susie L.1,MacKenzie Todd A.2,Klingensmith Georgeanna J.1,Murtfeldt Robert3,Garg Satish K.1

Affiliation:

1. From the Department of Pediatrics, Barbara Davis Center for Childhood Diabetes, Denver, Colorado;

2. Preventive Medicine and Biometrics, University of Colorado Health Sciences Center, Denver, Colorado; and

3. MiniMed, Sylmar, California.

Abstract

Purpose. To determine whether the use of continuous subcutaneous glucose monitoring will help in detecting unrecognized nocturnal hypoglycemia and in lowering hemoglobin A1c (HbA1c) levels (without increasing the risk for severe hypoglycemia) in children with type 1 diabetes. Methods. Eleven children with type 1 diabetes and HbA1c values consistently >8.0% were randomized either to the Continuous Glucose Monitoring System (CGMS) group or to the control group. The CGMS group used 6 3-day sensors within a 30-day period. Both groups self-monitored their blood glucose levels a minimum of 4 times daily. HbA1c levels were measured at the start, at 1-month, and after 3 months of study. Results. The 5 children using the CGMS had 17 asymptomatic episodes (85%) of glucose levels below 60 mg/dL (3.25 mmol/L) and 3 symptomatic episodes (15%) during the night in the study month. The 6 control children had 4 symptomatic nocturnal low episodes during the month. After the 30-day period of wearing the CGMS, the 5 children had a significantly lower mean HbA1c value compared with their initial value (mean ± standard error of the mean [SEM] decrease = .36% ± .07%). The mean decrease for the controls was .2% ± .2%. After 3 months, 4 of the 5 children who used the CGMS continued to have lower HbA1c values in comparison to their initial values (mean ± SEM decrease = 1.04% ± .43%). Three of the 6 control participants also had lower HbA1c values at 3 months (mean ± SEM decrease for the group = .62% ± .44%). No severe hypoglycemic events occurred in either the CGMS or the control groups. Conclusion. In this pilot trial, continuous subcutaneous glucose monitoring was helpful in detecting asymptomatic nocturnal hypoglycemia as well as in lowering HbA1c values without increasing the risk for severe hypoglycemia in children with type 1 diabetes.

Publisher

American Academy of Pediatrics (AAP)

Subject

Pediatrics, Perinatology and Child Health

Cited by 175 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3