Compliance With Alarm Limits for Pulse Oximetry in Very Preterm Infants

Author:

Clucas Luisa1,Doyle Lex W.1234,Dawson Jennifer3,Donath Susan15,Davis Peter G.13

Affiliation:

1. Departments of Obstetrics and Gynaecology

2. Paediatrics, University of Melbourne, Melbourne, Australia

3. Division of Newborn Services, Royal Women's Hospital, Carlton, Australia

4. Murdoch Children's Research Institute, Royal Children's Hospital, Melbourne, Australia

5. Clinical Epidemiology and Biostatistics Unit

Abstract

OBJECTIVE. The objective of this study was to determine the rate of compliance with hospital guidelines for alarm limits for pulse oximetry in preterm infants on oxygen therapy. METHODS. All infants admitted to the nurseries at the Royal Women's Hospital, Melbourne, Australia, with gestational age <32 weeks or birth weight <1500 g between August 2005 and February 2006 were eligible for inclusion. Data on the alarm limits set for infants on oxygen therapy were collected prospectively. The target saturation range recommended in written hospital guidelines was 88% to 92%, with alarm limits set at 85% and 94%. RESULTS. Data were prospectively collected for 144 subjects with mean (SD) gestational age 29.3 (2.4) weeks and birth weight 1226 (354) g; 1073 alarm limits were collected when infants were on oxygen. The lower alarm limit was set correctly 91.1% of the time. In contrast, the upper alarm limit was set correctly only 23.3% of the time: 76.5% of the time it was too high, and 23.8% of the time it was set at 100%. Infants with an upper alarm limit set correctly on a particular day had a significantly lower birth weight, gestational age, postmenstrual age, and postnatal age than infants who had the upper alarm limit set too high. Use of assisted ventilation, higher inspired oxygen concentrations, and more frequent changes in inspired oxygen concentration were all associated with improved odds of having an appropriately set upper alarm limit. CONCLUSIONS. This study suggests that current guidelines regarding the upper pulse oximeter alarm limit for infants receiving oxygen might be commonly exceeded, although compliance might be better for infants at higher risk of adverse outcomes. However, there might be less variation from guidelines for the lower alarm limit.

Publisher

American Academy of Pediatrics (AAP)

Subject

Pediatrics, Perinatology and Child Health

Cited by 64 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3