View-Synchronized Robotic Image-Guided Therapy for Atrial Fibrillation Ablation

Author:

Reddy Vivek Y.1,Neuzil Petr1,Malchano Zachary J.1,Vijaykumar Ragu1,Cury Ricardo1,Abbara Suhny1,Weichet Jiri1,McPherson Christina D.1,Ruskin Jeremy N.1

Affiliation:

1. From the Cardiac Arrhythmia Service (V.Y.R., Z.J.M., R.V., C.D.M., J.N.R.) and Radiology Department (R.C., S.A.), Massachusetts General Hospital and Harvard Medical School, Boston, and Homolka Hospital (P.N., J.W.), Prague, Czech Republic.

Abstract

Background— A robotic catheter navigation system has been developed that provides a significant degree of freedom of catheter movement. This study examines the feasibility of synchronizing this robotic navigation system with electroanatomic mapping and 3-dimensional computed tomography imaging to perform view-synchronized left atrial (LA) ablation. Methods and Results— This study consisted of a porcine experimental validation phase (9 animals) and a clinical feasibility phase (9 atrial fibrillation patients). Preprocedural computed tomography images were reconstructed to provide 3-dimensional surface models of the LA pulmonary veins and aorta. Aortic electroanatomic mapping was performed manually, followed by registration with the corresponding computed tomography aorta image using custom software. The mapping catheter was remotely manipulated with the robotic navigation system within the registered computed tomography image of the LA pulmonary veins. The point-to-surface error between the LA electroanatomic mapping data and the computed tomography image was 2.1±0.7 and 1.6±0.1 mm in the preclinical and clinical studies, respectively. The catheter was remotely navigated into all pulmonary veins, the LA appendage, and circumferentially along the mitral valve annulus. In 7 of 9 animals, circumferential radiofrequency ablation lesions were applied periostially to ablate 11 pulmonary veins. In patients, all of the pulmonary veins were remotely electrically isolated in an extraostial fashion. Adjunctive ablation included superior vena cava isolation in 6 patients, cavotricuspid isthmus ablation in 5 patients, and ablation of sites of complex fractionated activity and atypical LA flutters in 3 patients. Conclusions— This study demonstrates the safety and feasibility of an emerging paradigm for atrial fibrillation ablation involving the confluence of 3 technologies: 3-dimensional imaging, electroanatomic mapping, and remote robotic navigation.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 81 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3