Adverse Outcomes of Interrupted Precordial Compression During Automated Defibrillation

Author:

Yu Ting1,Weil Max Harry1,Tang Wanchun1,Sun Shijie1,Klouche Kada1,Povoas Heitor1,Bisera Joe1

Affiliation:

1. From The Institute of Critical Care Medicine, Palm Springs (T.Y., M.H.W., W.T., S.S., K.K., H.P., J.B.), and the Keck School of Medicine of the University of Southern California, Los Angeles (M.H.W., W.T., S.S., J.B.), Calif.

Abstract

Background Current versions of automated external defibrillators (AEDs) require frequent stopping of chest compression for rhythm analyses and capacity charging. The present study was undertaken to evaluate the effects of these interruptions during the operation of AEDs. Methods and Results Ventricular fibrillation was electrically induced in 20 male domestic swine weighing between 37.5 and 43 kg that were untreated for 7 minutes before CPR was started. Defibrillation was attempted with up to 3 sequential 150-J biphasic shocks, but each was preceded by 3-, 10-, 15-, or 20-second interruptions of chest compression. The interruptions corresponded to those that were mandated by commercially marketed AEDs for rhythm analyses and capacitor charge. The sequence of up to 3 electrical shocks and delays were repeated at 1-minute intervals until the animals were successfully resuscitated or for a total of 15 minutes. Spontaneous circulation was restored in each of 5 animals in which precordial compression was delayed for 3 seconds before the delivery of the first and subsequent shocks but in none of the animals in which the delay was >15 seconds before the delivery of the first and subsequent shocks. Longer intervals of CPR interventions were required, and there was correspondingly greater failure of resuscitation in close relationship to increasing delays. The durations of interruptions were inversely related to the durations of subthreshold levels of coronary perfusion pressure. Postresuscitation arterial pressure and left ventricular ejection fraction were more severely impaired with increasing delays. Conclusions Interruptions of precordial compression for rhythm analyses that exceed 15 seconds before each shock compromise the outcome of CPR and increase the severity of postresuscitation myocardial dysfunction.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3