Localized kinking during deployment of a flow redirection lumen device (FRED) could be due to excessive pushing

Author:

Beppu Mikiya1,Kuramoto Yoji1,Abe Soichiro1,Namitome Satoshi1,Yoshimura Shinichi1

Affiliation:

1. Department of Neurosurgery, Hyogo College of Medicine, Nishinomiya, Hyogo, Japan.

Abstract

Background: The safety and efficacy analysis of flow redirection lumen device (FRED) demonstrated the excellent safety profile of FREDs for aneurysm treatment. We describe the first case in which FRED deployment for a paraclinoid aneurysm resulted in in-stent stenosis, necessitating balloon angioplasty, and an additional stent. Case Description: A 50-year-old woman had a left paraclinoid aneurysm with a maximum diameter of 6.1 mm. We planned FRED deployment. We experienced in-stent stenosis just after the deployment of a FRED. Devices such as guidewires and catheters could not cross the lesion through the FRED because of an obstruction in the FRED. Balloon angioplasty and subsequent stenting resolved thrombosis and kinking. FRED has potential for kinking locally. Conclusion: Surgeons should consider this possibility when treating cerebral aneurysm using FRED. Cone-beam computed tomography after deployment of FRED may be useful for evaluating the stent shape.

Publisher

Scientific Scholar

Subject

Neurology (clinical),Surgery

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. A Simulation Framework for Magnetic Continuum Robots;IEEE Robotics and Automation Letters;2022-07

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3