Abstract
The Sensititre broth microdilution antibiotic susceptibility testing system was found to be comparable to a reference broth dilution method. Of 3,194 minimum inhibitory concentration data pairs, 93.9% were within the acceptable +/- 1-log2 dilution range. Only rare (0.6%) very major errors were found between methods.
Publisher
American Society for Microbiology
Reference7 articles.
1. Evaluation of the Micro-Media System for quantitative antimicrobial drug susceptibility testing: a collaborative study;Barry A. L.;Antimicrob. Agents Chemother.,1978
2. The replicator method for identification and biotyping of common bacterial isolates;Fuchs P. C.;Lab. Med.,1976
3. Gavan T. L. and D. A. Butler. 1974. An automated microdilution method for antimicrobial susceptibility testing p. 88-93. In A. Balows (ed.) Current techniques for antibiotic susceptibility testing. Charles C Thomas Publisher Springfield Ill.
4. Evaluation of the Sensititre system for quantitative antimicrobial drug susceptibility testing: a collaborative study;Gavan T. L.;Antimicrob. Agents Chemother.,1980
5. A microdilution method for antibiotic susceptibility testing: an evaluation;Gavan T. L.;Am. J. Clin. Pathol.,1970