Accuracy of Results Obtained by Performing a Second Ligase Chain Reaction Assay and PCR Analysis on Urine Samples with Positive or Near-Cutoff Results in the LCx Test for Chlamydia trachomatis

Author:

Castriciano S.1,Luinstra K.12,Jang D.12,Patel J.3,Mahony J.142,Kapala J.3,Chernesky M.142

Affiliation:

1. Hamilton Regional Laboratory Medicine Program

2. the Father Sean O'Sullivan Research Centre, St. Joseph's Healthcare, McMaster University, Hamilton, Ontario, Canada L8N 4A6

3. Department of Microbiology, Gamma-Dynacare Medical Laboratories, Brampton, Ontario, Canada L6T 5M3

4. Departments of Pathology and Molecular Medicine

Abstract

ABSTRACT Nucleic acid amplification assays such as the ligase chain reaction and PCR have encountered reproducibility problems. The initial extract and a newly extracted aliquot of urine specimens ( n = 120) which had signal-to-cutoff (S/CO) ratios above 0.80 by the LCx Chlamydia assay were retested. Nucleic acid was extracted from an additional urine sample for testing by the AMPLICOR PCR Chlamydia assay. Fifteen percent (18 of 120) of the urine specimens were negative by all repeat tests (initial mean S/CO ratio by the LCx Chlamydia assay, 0.93; S/CO ratio range, 0.80 to 3.30). Repeat testing of the 102 specimens with possible positive results by the LCx Chlamydia assay by use of the initially extracted aliquot confirmed the results for 95 (93.1%) of the specimens; repeat testing of a newly extracted aliquot confirmed the results for 87 (85.3%) of the specimens. Twenty specimens had discordant results by the two repeat LCx Chlamydia assays. A total of 78 of 102 (76.5%) of the specimens were positive by the AMPLICOR PCR, and the AMPLICOR PCR confirmed the results for 82.1% (78 of 95) and 89.6% (78 of 87) of the specimens positive by the two repeat LCx Chlamydia assays, respectively. Some of the discrepancies observed by multiple repeat tests may have been due to specimen mislabeling or contamination during performance of the procedure rather than to the LCx Chlamydia assay. Both assays suffered from a lack of reproducibility on repeat testing with a small proportion of specimens, probably due to the presence of low levels of DNA, the presence of variable amounts of amplification inhibitors, and the loss of DNA during extraction.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Cited by 24 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Advances in ligase chain reaction and ligation-based amplifications for genotyping assays: Detection and applications;Mutation Research/Reviews in Mutation Research;2017-07

2. Utilization Management in a Large Community Hospital;Utilization Management in the Clinical Laboratory and Other Ancillary Services;2016-12-01

3. Evaluation of Copan FLOQSwab for the molecular detection ofChlamydia trachomatisby Abbott RealTime CT PCR;Acta Clinica Belgica;2015-06-24

4. Outreach Implementation Requirements: A Case Study;Clinical Laboratory Management;2014-10-30

5. Advances in sampling and screening for chlamydia;Future Microbiology;2013-03

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3