Comparison of Two Amplification Technologies for Detection and Quantitation of Human Immunodeficiency Virus Type 1 RNA in the Female Genital Tract

Author:

Bremer James1,Nowicki Marek2,Beckner Suzanne3,Brambilla Donald4,Cronin Mike5,Herman Steven6,Kovacs Andrea2,Reichelderfer Patricia7

Affiliation:

1. Rush Medical College, Chicago, Illinois 606121;

2. University of Southern California, Los Angeles, California 900332;

3. Westat, Rockville, Maryland 208503;

4. New England Research Institutes, Watertown, Massachusetts 024724;

5. Organon Teknika Corporation, Durham, North Carolina 277125;

6. Department of Infectious Diseases, Roche Molecular Systems, Inc., Alameda, California 945016; and

7. National Institute of Child Health and Human Development, Bethesda, Maryland 208927

Abstract

ABSTRACT Human immunodeficiency virus type 1 (HIV-1) RNA levels in female genital tract and peripheral blood samples were compared using two commercial amplification technologies: the Roche AMPLICOR HIV-1 MONITOR test and either the Organon Teknika nucleic acid sequence-based amplification (NASBA-QT) assay or the NucliSens assay. Estimates of HIV-1 RNA copy number were derived from internal kit standards and analyzed unadjusted and adjusted to a common set of external standards. We found a discordance rate of approximately 18% between the two technologies for the detection of HIV-1 in either the genital tract or peripheral blood samples. Detection discordance was not consistent among specimens or among women. There were no significant differences in adjusted or unadjusted estimates of HIV-1 RNA copy number in the genital tract samples using the AMPLICOR HIV-1 MONITOR test and either the NASBA-QT assay or the NucliSens assay. In addition, the estimated HIV-1 RNA copy number in peripheral blood samples did not differ when tested with the NucliSens assay and the AMPLICOR HIV-1 MONITOR test using kit standards. However, there was a significant difference in estimated RNA copy number between the NASBA-QT assay and the AMPLICOR HIV-1 MONITOR test for internal kit standards, which, as we have previously shown, was eliminated after adjustment with the external standards. Our results suggest that the Roche and Organon Teknika assays are equivalent for quantifying HIV-1 RNA in female genital tract specimens, although variation in detection does exist.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Cited by 26 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3