ABSTRACT
We tracked and documented the time of positivity of blood cultures by using the BACTEC 9120 (Becton Dickinson Diagnostic Instrument Systems) blood culture system over a 5-year study period. A 7-day protocol of the incubation period was selected, and a total of 11,156 blood cultures were evaluated. The clinically significant microorganisms (32.95%) were isolated in 3,676 specimens. Gram-positive and -negative bacterial isolation rates were found to be 41.07 and 44.88%, respectively. Yeasts were found in 14.03% of all pathogens. Both the false-positivity and -negativity rates were very low (0.1 and 0.3%, respectively). The mean detection times for all of the pathogens were determined to be 19.45 h. Yeasts, nonfermentative gram-negative bacteria, and
Brucella melitensis
strains were isolated within 5 days. By taking these data into account, we decided to establish a 5-day-incubation protocol in our laboratory instead of the 7 days that are commonly used.
Publisher
American Society for Microbiology
Reference17 articles.
1. Rapid diagnosis of Brucella bacteremia by using the BACTEC 9240 system
2. Bayram, A., G. N. Alkan, and I·. Balcı. 1997. Microorganisms isolated from a continuous monitoring blood culture system: BACTEC 9240. J. Turk. Microbiol. Soc.27:64-67.
3. Clinical comparison of BACTEC 9240 plus aerobic/F resin bottles and the isolator aerobic culture system for detection of bloodstream infections
4. Gedikoglu, S., S. Helvacı, C. Özakın, F. Gökırmak, and K. Kılıçturgay. 1996. Detection of Brucella melitensis by BACTEC NR 730 and BACTEC 9120 systems. Eur. J. Epidemiol.12:649-650.
5. Gray, J., M. Brockwell, and L. Das. 1998. Experience of changing between signal and BACTEC 9240 blood culture systems in a children's hospital. J. Clin. Pathol.51:302-305.
Cited by
49 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献