Affiliation:
1. Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, Minnesota 55905, USA. Cockerill.franklin@mayo.edu
Abstract
The Plus Aerobic/F resin bottle of the BACTEC 9240 automated blood culture system (Becton Dickinson Diagnostic Instrument Systems, Sparks, Md.) was compared with aerobic culture of the Isolator system (Wampole Laboratories, Cranbury, N.J.) for the detection of bloodstream microorganisms from 6,145 blood cultures collected from adult patients with suspected septicemia. The BACTEC resin bottles were incubated for 7 days, and the sediment from the Isolator tube was inoculated to sheep blood and chocolate agars which were incubated for 72 h and to inhibitory mold, brain heart infusion, and Sabouraud agars which were incubated for 21 days. A total of 622 microorganisms were recovered from 583 blood cultures. The BACTEC resin bottle recovered statistically significantly more pathogens overall than the Isolator system (P = 0.0006). When individual pathogens isolated from either system for a 7-day study period were assessed, it was determined that the BACTEC resin bottle detected statistically significantly more isolates of Staphylococcus aureus (P = 0.0113) and coagulase-negative Staphylococcus spp. (P = 0.0029) than the Isolator system. The BACTEC resin bottle also detected statistically significantly more bloodstream infections (septic episodes) caused by coagulase-negative Staphylococcus spp. (P = 0.0146). The Isolator system recovered statistically significantly more contaminants overall (P < 0.0001), and among this group of microorganisms, recovered statistically significantly more Bacillus spp. (P < 0.0001), coagulase-negative Staphylococcus spp. (P < 0.0001), and viridans group Streptococcus spp. (P = 0.0156). The Isolator system detected statistically significantly more isolates of Histoplasma capsulatum (P = 0.004), but all of these isolates were detected at > or = 7 days of incubation of fungal plates, i.e., after the system to system comparison study period (7 days). In blood culture sets which produced growth of the same pathogen in both systems, there was a statistically significant difference in median time to detection for all pathogens combined favoring the BACTEC resin bottle over the Isolator tube (P < 0.05). When assessing individual microorganisms, the median times for detection of S. aureus, Enterococcus spp., and Pseudomonas spp. were all statistically significantly less for the BACTEC system (P < 0.05). The BACTEC instrument had 79 (1.3%) false positive signals. The BACTEC system required less processing time than the Isolator system and eliminates the hands-on time for detection of positive cultures required with the Isolator system.
Publisher
American Society for Microbiology
Cited by
39 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献