Two-Center Collaborative Evaluation of Performance of the BD Phoenix Automated Microbiology System for Identification and Antimicrobial Susceptibility Testing of Gram-Negative Bacteria

Author:

Menozzi Maria Grazia1,Eigner Ulrich2,Covan Silvia1,Rossi Sabina1,Somenzi Pietro1,Dettori Giuseppe1,Chezzi Carlo1,Fahr Anne-Marie2

Affiliation:

1. Section of Microbiology, Department of Pathology and Laboratory Medicine, University of Parma, I-43100 Parma, Italy

2. Department of Microbiology, Laboratory Group Heidelberg, D-69126 Heidelberg, Germany

Abstract

ABSTRACT The performance of the BD Phoenix Automated Microbiology System (BD Diagnostic Systems, Sparks, MD) was assessed for identification (ID) and antimicrobial susceptibility testing (AST) of the majority of clinically encountered bacterial isolates in a European collaborative two-center trial. A total of 494 bacterial isolates including various species of the Enterobacteriaceae and 110 nonfermentative gram-negative bacteria were investigated: of these, 385 were single patient isolates, and 109 were challenge strains tested at one center. The performance of the Phoenix extended-spectrum β-lactamase (ESBL) test was also evaluated for 203 strains of Escherichia coli , Klebsiella pneumoniae , and Klebsiella oxytoca included in the study. Forty-two antimicrobial drugs were tested, including members of the following drug classes: aminoglycosides, β-lactam antibiotics, β-lactam/β-lactamase inhibitors, carbapenems, cephems, monobactams, folate antagonists, quinolones, and others. Phoenix system ID results were compared to those of the laboratories' routine ID systems (API 20E and API CHE, ATB ID32E, ID32GN, and VITEK 2 [bioMérieux, Marcy l'Etoile, France]); Phoenix AST results were compared to those of frozen standard broth microdilution (SBM) panels according to NCCLS (now CLSI) guidelines (NCCLS document M100-S9, approved standard M7-A4). Discrepant results were repeated in duplicate. Concordant IDs of 98.4 and 99.1% were observed for the Enterobacteriaceae and the nonfermentative group, respectively. For AST results, the overall essential agreement was 94.2%; the category agreement was 97.3%; and the very major error rate, major error rate, and minor error rate were 1.6, 0.6, and 1.9%, respectively. In terms of ESBL detection, Phoenix results were 98.5% concordant with those of the reference system, with 98.0% sensitivity and 98.7% specificity. In conclusion, the Phoenix ID results showed high agreement with results of the systems to which they were being compared: the AST performance was highly equivalent to that of the SBM reference method, and the system proved to be very accurate for the detection of ESBL producers.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司