Evaluation of the BD Phoenix Automated Microbiology System for Identification and Antimicrobial Susceptibility Testing of Enterobacteriaceae

Author:

Carroll Karen C.12,Glanz Brian D.2,Borek Anita P.2,Burger Chad2,Bhally Hasan S.12,Henciak Susan2,Flayhart Diane2

Affiliation:

1. Division of Medical Microbiology

2. Microbiology Laboratory, The Johns Hopkins Medical Institutions, Baltimore, Maryland

Abstract

ABSTRACT We evaluated the accuracy of the BD Phoenix system for the identification (ID) and antimicrobial susceptibility testing (AST) of 251 isolates of the family Enterobacteriaceae representing 31 species. Organisms were inoculated onto the Phoenix panel according to the manufacturer's instructions. The results from conventional biochemical tests were used for the reference method for ID. Agar dilution, performed according to the CLSI guidelines, was the reference AST method. Essential and categorical agreements were determined. The overall levels of agreement for the genus- and species-level identifications were 95.6% and 94.4%, respectively. Fourteen isolates were incorrectly identified by the Phoenix system; 10 of these were incorrectly identified to the species level. Three of these were Enterobacter ( Pantoea ) species and four of these were Shigella spp. misidentified as Escherichia coli . For AST results, the essential and categorical agreements were 98.7% and 97.9%, respectively. The very major error, major error, and minor error rates were 0.38%, 0.33%, and 1.8%, respectively. Six isolates (three E. coli isolates and three Klebsiella isolates) were extended-spectrum β-lactamase producers. All six were flagged by the Phoenix system expert rules. The Phoenix system compares favorably to traditional methods for ID and AST of Enterobacteriaceae .

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference9 articles.

1. Evaluation of the Automated Phoenix System for Potential Routine Use in the Clinical Microbiology Laboratory

2. Analysis of the Comparative Workflow and Performance Characteristics of the VITEK 2 and Phoenix Systems

3. Farmer, J. J., III. 2003. Enterobacteriaceae: introduction and identification, p. 636-653. In P. R. Murray E. J. Baron, J. H. Jorgensen, M. A. Pfaller, and R. H. Yolken (ed.), Manual of clinical microbiology, 8th ed. American Society for Microbiology, Washington, D.C.

4. Evaluation of the Etest ESBL and the BD Phoenix, VITEK 1, and VITEK 2 Automated Instruments for Detection of Extended-Spectrum Beta-Lactamases in Multiresistant Escherichia coli and Klebsiella spp

5. National Committee for Clinical Laboratory Standards. 2004. Performance standards for antimicrobial susceptibility testing; fourteenth informational supplement. M100-S14. National Committee for Clinical Laboratory Standards Wayne Pa.