Abstract

The Vitek AutoMicrobic System with GSC-plus cards and the Abbott MS-2 system were tested in parallel and the results were compared directly with those of a reference microdilution minimal inhibitory concentration (MIC) procedure on a group of 262 clinical isolates of the family Enterobacteriaceae and of Pseudomonas aeruginosa. Results of both systems were compared with the reference MIC for category agreement, and in addition, the Vitek MICs were compared with those obtained by the reference procedure. The Vitek system provided an essential category correlation of 89.4% for enteric bacteria and 97.0% for P. aeruginosa. Vitek MICs agreed within 1 twofold dilutional increment for 86.3% of the enteric bacteria tested and for 96.2% of the P. aeruginosa isolates. The Abbott MS-2 essential categoric agreement was 92.0% for enteric bacteria and 92.4% for P. aeruginosa. If only aminoglycosides or carbenicillin were considered for P. aeruginosa isolates, the essential category agreement was 92.5% for the Vitek and 93.3% for the MS-2. The majority of MS-2 category errors (13 of 19) with P. aeruginosa involved gentamicin results on isolates whose reference MICs were 8 micrograms/ml and whose MS-2 results were susceptible (MIC less than or equal to 4 micrograms/ml). Retesting of the P. aeruginosa isolates in calcium-supplemented MS-2 broth increased the essential agreement for the aminoglycosides to 97.5%.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference16 articles.

1. Automated microbiological detection/identification system;Aldridge C.;J. Clin. Microbiol.,1977

2. Clinical evaluation of automated antibiotic susceptibility testing with the MS-2 system;Barnes W. G.;J. Clin. Microbiol.,1980

3. Antibiotic susceptibility by a standardized single disk method;Bauer A. W.;Am. J. Clin. Pathol.,1966

4. Gavan T. L. and A. L. Barry. 1980. Microdilution test procedures p. 459-462. In E. H. Lennette A. Balows W. J. Hausler Jr. and J. P. Truant (ed.). Manual of clinical microbiology 3rd ed. American Society for Microbiology Washington D.C.

5. Gavan T. L. and D. A. Butler. 1974. An automated microdilution method for antimicrobic susceptibility testing p. 88-93. In A. Balows (ed.) Current techniques for antibiotic susceptibility testing. Charles C Thomas Publisher Springfield Ill.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3