Affiliation:
1. Service de Microbiologie, Hôpital Saint Louis, Paris
2. Service de Microbiologie, Hôpital Paul Brousse, Villejuif, France
Abstract
ABSTRACT
A comparative study was designed to evaluate the identification (ID) and antimicrobial susceptibility testing (AST) performances of the BD Phoenix Automated Microbiology System (Becton Dickinson Diagnostic Systems [BD], Pont de Claix, France). A total of 305 single clinical isolates were collected, and comparisons were made with routine manual methods in use in our microbiology laboratories. The percentages of correct IDs were 93.3, 89.4, 91.8, and 85.7% for enterobacteria, nonfermenting gram-negative bacilli, staphylococci, and streptococci-enterococci, respectively. The median ID time was 3 h, and the median time for AST was 10 h 30 min. AST results showed variable percentages of errors for the different antibiotics. None of the enterobacteria and 0.3% of
Pseudomonas aeruginosa
isolates showed a very major error (VME). Only one strain of
Staphylococcus aureus
showed a VME with oxacillin. We demonstrate here the efficiency of the Phoenix system, which can be used for the majority of strains encountered in a university-based laboratory, for ID and AST.
Publisher
American Society for Microbiology
Cited by
86 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献