Abstract
ABSTRACT
The Abbott RealTime human immunodeficiency virus type 1 (HIV-1) assay (ART) and the Cobas AmpliPrep/Cobas TaqMan HIV-1 test (CTM) are commercially available assays for quantification of HIV-1 RNA in plasma. We evaluated performance characteristics, workflow, throughput, reliability, and direct costs of these assays. Both assays yielded good correlation of quantitative results (
r
= 0.95) among clinical specimens, with a mean difference of −0.34 log
10
copies/ml. Testing of healthy donor plasma specimens yielded “target not detected” results by ART, with “HIV-1 RNA detected, <40 copies/ml” results for 3.3% (3 of 90 samples) of these specimens by CTM. Both the m2000sp/m2000rt (ART) and docked CAP/CTM96 (CTM) instrument systems were capable of operating with continuous, uninterrupted workflow. When daily maintenance and cleaning were included, ART and CTM run durations (5 h 52 min and 6 h 4 min, respectively) and hands-on times (53 min and 46 min, respectively) were similar for a run batch size of 24. While ART was more flexible in terms of run batch size, CTM required fewer user interventions and consistently produced higher specimen throughput rates at 8, 16, and 24 h. Assay run failure rates were 6.3% (1 of 16 runs) and 4.2% (1 of 24 runs) for ART and CTM, respectively (
P
= 1.000), with invalid specimen result rates of 1.0% (5 of 495 specimens) and 2.8% (11 of 399 specimens), respectively (
P
= 0.073). Direct reagent and consumable costs for each assay were comparable (difference of <10%). In selecting an assay for implementation, laboratories should consider how various assay and instrument features might impact laboratory operation and patient care.
Publisher
American Society for Microbiology
Reference24 articles.
1. Bland, J. M., and D. G. Altman. 1999. Measuring agreement in method comparison studies. Stat. Methods Med. Res.8:135-160.
2. Braun, P., R. Ehret, F. Wiesmann, F. Zabbai, M. Knickmann, R. Kuhn, S. Thamm, G. Warnat, and H. Knechten. 2007. Comparison of four commercial quantitative HIV-1 assays for viral load monitoring in clinical daily routine. Clin. Chem. Lab. Med.45:93-99.
3. Human Immunodeficiency Virus Type 1 (HIV-1) Plasma Load Discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR Version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 Assays
4. Application OF Branched DNA Signal Amplification to Monitor Human Immunodeficiency Virus Type 1 Burden in Human Plasma
5. Probit analysis 1971
Cited by
39 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献