Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya

Author:

Mwau MatiluORCID,Schaffer Sven,Kimani Humphrey,Kasiano Purity,Ogolla Francis,Ajema Elizabeth,Adoyo Scriven,Nyairo Ednah,Saleri Norah,Nair Sangeetha Vijaysri

Abstract

Introduction HIV-1 viral Load (VL) testing is recommended for the monitoring of antiretroviral treatment. Dried Blood Spots (DBS) are an effective sample type in resource limited settings, where safe phlebotomy and reliable shipping are hard to guarantee. In HIV high burden countries, high throughput assays can improve access to testing services. The Hologic Aptima HIV-1 Quant Dx Assay (Aptima Assay) is a high throughput assay that runs on the CE-IVD approved Panther platform. The objectives of this study were to assess the performance characteristics of Aptima for VL monitoring using plasma and venous DBS specimens and to determine the stability of HIV-1 RNA in DBS. Materials and methods This was a cross-sectional study of 2227 HIV infected adults visiting health facilities in Nairobi and Busia, Kenya. Each provided a venous blood sample; plasma was prepared from 1312 samples while paired DBS samples and plasma were prepared from the remaining 915 samples. The agreement between the Aptima assay and the Abbott RealTime HIV-1 Assay (Abbott RT) was analysed by comparing the HIV-1 VL in both assays at the medical decision point of 1000 copies/mL. To assess stability of HIV-1 RNA in DBS, VL in DBS spotted on day 0 were compared with that from the same DBS card after 21 days of storage at room temperature. Results Overall, 436 plasma samples had quantifiable results in both Aptima and Abbott RT. The agreement between the two assays at 1000 copies/mL was 97.48% with a Pearson’s correlation coefficient (r) of 0.9589 and gave a mean bias of 0.33 log copies/mL on Bland-Altman analysis. For fresh DBS, the agreement in both assays was 94.64% at 1000 copies/mL, with an r of 0.8692 and a mean bias of 0.35 log copies/mL. The overall agreement between DBS tested in Aptima on day 0 versus day 21 was 95.71%, with a mean bias of -0.154. Conclusion The Aptima HIV-1 Quant Dx assay is an accurate test for VL monitoring of HIV-1 using DBS and plasma sample types in Kenya.

Funder

Hologic, Inc

Publisher

Public Library of Science (PLoS)

Subject

Multidisciplinary

Reference30 articles.

1. UNAIDS HIV and AIDS Statistics 2021 https://www.unaids.org/en/resources/fact-sheet

2. HIV/AIDS. 90-90-90: an ambitious treatment target to help end the AIDS epidemic. Joint United Nations Programme on HIV/AIDS, Geneva, Switzerland. 2014.

3. Predicting treatment failure in adults and children on antiretroviral therapy: a systematic review of the performance characteristics of the 2010 WHO immunologic and clinical criteria for virologic failure;GW Rutherford;AIDS,2014

4. Virological follow-up of adult patients in antiretroviral treatment programmes in sub-Saharan Africa: a systematic review;RE Barth;Lancet Infect Dis,2010

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3