Performance of the New Bayer VERSANT HCV RNA 3.0 Assay for Quantitation of Hepatitis C Virus RNA in Plasma and Serum: Conversion to International Units and Comparison with the Roche COBAS Amplicor HCV Monitor, Version 2.0, Assay

Author:

Beld Marcel1,Sentjens Roel2,Rebers Sjoerd1,Weegink Christine2,Weel Jan1,Sol Cees1,Boom René1

Affiliation:

1. Section of Clinical Virology, Department of Medical Microbiology

2. Section of Hepatology, Department of Gastro-Enterology, Academic Medical Center, 1105 AZ Amsterdam, The Netherlands

Abstract

ABSTRACT We have evaluated the VERSANT HCV RNA 3.0. Assay (HCV 3.0 bDNA assay) (Bayer Diagnostics, Berkeley, Calif.), which is an improved signal amplification procedure for the HCV 2.0 bDNA assay for the quantitation of hepatitis C virus (HCV) RNA in serum or plasma of HCV-infected individuals. The HCV 3.0 bDNA assay has a linear dynamic range of 2.5 × 10 3 to 4.0 × 10 7 HCV RNA copies per ml (c/ml). The performance of the HCV 3.0 bDNA assay was evaluated using three different test panels. An overall specificity of 96.8% relative to the detection limit of the HCV 3.0 bDNA assay was found. The intra- and interrun reproducibilities for both the dilution panel and the NAP (AcroMetrix, Benicia, Calif.) panel were consistent with coefficients of variation of less than 9%. Quantitation with the HCV 3.0 bDNA assay was linear over the entire range of both panels (ranges of 4.4 × 10 3 to 3.5 × 10 6 c/ml and 5 × 10 3 to 2 × 10 6 IU/ml, respectively), with correlation coefficients of 0.999, slopes close to one, and intercepts close to zero. The regression equation indicated that 1 IU corresponded to about 4.8 copies of HCV RNA. A correlation coefficient of 0.941 was found for HCV RNA values (in international units per milliliter) obtained from the HCV 3.0 bDNA assay and the HCV Monitor version 2.0 assay (HCV Monitor 2.0 assay) (Roche Diagnostic Systems, Branchburg, N.J.). Quantitative results obtained close to the lower limit of the HCV 3.0 bDNA assay might imply that its lower limit should be reconsidered and raised, if necessary. It appeared that quantitation values obtained from the HCV Monitor 2.0 assay of between 5 × 10 2 and 10 5 IU/ml were in general higher than those obtained from the HCV 3.0 bDNA assay, whereas values obtained from the HCV Monitor 2.0 assay were underestimated for samples with HCV RNA levels above 10 5 IU/ml.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3