Affiliation:
1. Department of Microbiology and Immunology, University of Washington, School of Medicine, Seattle, Washington 98195
Abstract
Electrical impedance changes in the medium were studied during traditional broth dilution tests. Tests involved clinical isolates of
Escherichia coli, Staphylococcus aureus, Pseudomonas aeruginosa, Enterobacter
sp.,
Klebsiella
, and enterococcus and ten antibiotics. Minimal inhibitory concentration (MIC) values from overnight visual readings were compared with MIC values determined from electrical impedance end points, using inocula of 10
6
organisms per ml. Ninety-three percent of the results were within one twofold dilution of each other. When the impedance end point was determined at 6 h, this correlation was lowered to 34%. By increasing the initial electrical impedance inoculum 10-fold, the correlation between the 6-h impedance MIC and the overnight visual MIC was improved to 74%. Ampicillin, tetracycline, and polymyxin E results accounted for most discrepancies. Continuous monitoring of impedance changes suggested that better correspondence could be obtained by adjusting the end point criteria for the 6-h impedance MIC with different antibiotics. Electrical impedance methods for reading bacterial end points in automated clinical laboratory instruments appear promising.
Publisher
American Society for Microbiology
Subject
Infectious Diseases,Pharmacology (medical),Pharmacology
Reference9 articles.
1. Cady P. 1975. Rapid automated bacterial identification by impedance measurement p. 73-99. In C. G. Heden and T. Illeni (ed.) New approaches to the identification of microorganisms. John Wiley & Sons Inc. New York.
2. Antibiotic sensitivity testing. Report of an international collaborative study;Ericson H. M.;Acta Pathol. Microbiol. Scand. Sect. B,1971
3. Prototype of a fully automated device for the determination of bacterial antibiotic susceptibility in the clinical laboratory;Isenberg H. P.;Appl. Microbiol.,1971
4. Rapid automated diagnosis of bacteremia by impedance detection;Kagan R. L.;J. Clin. Microbiol.,1977
5. Relationship of early readings of minimal inhibitory concentrations to the results of overnight tests;Lampe M. F.;Antimicrob. Agents Chemother.,1975