Affiliation:
1. University of California, San Diego 1 , 9500 Gilman Drive, La Jolla, California 92093-0417, USA
2. General Atomics 2 , P.O. Box 85608, San Diego, California 92186-5608, USA
Abstract
Two camera systems are installed on the DIII-D tokamak at the toroidal positions of 90° (90° system) and 225° (225° system), respectively. The cameras have two types of relay optics, namely, a coherent optical fiber bundle and a periscope system. The periscope system provides absolute intensity calibration stability while sacrificing resolution (10 lp/mm), while the fiber system provides high resolution (16 lp/mm) while sacrificing calibration stability. The periscope is available only for the 90° system. The optics of the 225° system were designed for view stability, repeatability, and easy maintenance. The cameras are located inside optimized neutron, x ray and magnetic shielding in order to reduce electronics damage, reboots, and magnetic and neutron interference, increasing the overall system reliability. An automated filter wheel, providing remote filter change, allows for remote wavelength selection. A software suite automates camera acquisition and data storage, allowing for remote operation and reduced operator involvement. System metadata is used to streamline the data analysis workflow, particularly for intensity calibration. The spatial calibration uses multiple observable wall features, resulting in a reconstruction accuracy ≤2 cm.
Funder
U.S. Department of Energy
Cited by
2 articles.