Affiliation:
1. Department of Physics and Astronomy, University of California Los Angeles, Los Angeles, California 90095, USA
Abstract
We have developed a non-collective Thomson scattering diagnostic for measurements of electron density and temperature on the Large Plasma Device. A triple grating spectrometer with a tunable notch filter is used to discriminate the faint scattering signal from the stray light. In this paper, we describe the diagnostic and its calibration via Raman scattering and present the first measurements performed with the fully commissioned system. Depending on the discharge conditions, the measured densities and temperatures range from 4.0 × 1012 to 2.8 × 1013 cm−3 and from 1.2 to 6.8 eV, respectively. The variation of the measurement error with plasma parameters and discharges averaged is also discussed.
Funder
U.S. Department of Energy
National Science Foundation
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献