Patient-based quality control for glucometers using the moving sum of positive patient results and moving average

Author:

Yee Lim Chun1,Badrick Tony2,Ping Loh Tze3

Affiliation:

1. Engineering Cluster, Singapore Institute of Technology, Singapore

2. RCPA Quality Assurance Programs, St Leonards, Sydney, Australia; International Federation of Clinical Chemistry and Laboratory Medicine Committee on Analytical Quality

3. Department of Laboratory Medicine, National University Hospital, Singapore

Abstract

Introduction: The capability of glucometer internal quality control (QC) in detecting varying magnitude of systematic error (bias), and the potential use of moving sum of positive results (MovSum) and moving average (MA) techniques as potential alternatives were evaluated. Materials and methods: The probability of error detection using routine QC and manufacturer’s control limits were investigated using historical data. Moving sum of positive results and MA algorithms were developed and optimized before being evaluated through numerical simulation for false positive rate and probability of error detection. Results: When the manufacturer’s default control limits (that are multiple times higher than the running standard deviation (SD) of the glucometer) was used, they had 0-75% probability of detecting small errors up to 0.8 mmol/L. However, the error detection capability improved to 20-100% when the running SD of the glucometer was used. At a binarization threshold of 6.2 mmol/L and block sizes of 200 to 400, MovSum has a 100% probability of detecting a bias that is greater than 0.5 mmol/L. Compared to MovSum, the MA technique had lower probability of bias detection, especially for smaller bias magnitudes; MA also had higher false positive rates. Conclusions: The MovSum technique is suited for detecting small, but clinically significant biases. Point of care QC should follow conventional practice by setting the control limits according to the running mean and SD to allow proper error detection. The glucometer manufacturers have an active role to play in liberalizing QC settings and also enhancing the middleware to facility patient-based QC practices.

Publisher

Croatian Society for Medical Biochemistry and Laboratory Medicine

Subject

Biochemistry, medical,Clinical Biochemistry

Cited by 13 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3