IntroductionAutoverification (AV) is a postanalytical tool that uses algorithms to validate test results according to specified criteria. The Clinical and Laboratory Standard Institute (CLSI) document for AV of clinical laboratory test result (AUTO-10A) includes recommendations for laboratories needing guidance on implementation of AV algorithms. The aim was to design and validate the AV algorithm for biochemical tests.
Materials and methodsCriteria were defined according to AUTO-10A. Three different approaches for algorithm were used as result limit checks, which are reference range, reference range ± total allowable error, and 2nd and 98th percentile values. To validate the algorithm, 720 cases in middleware were tested. For actual cases, 3,188,095 results and 194,520 reports in laboratory information system (LIS) were evaluated using the AV system. Cohen’s kappa (κ) was calculated to determine the degree of agreement between seven independent reviewers and the AV system.
ResultsThe AV passing rate was found between 77% and 85%. The highest rates of AV were in alanine transaminase (ALT), direct bilirubin (DBIL), and magnesium (Mg), which all had AV rates exceeding 85%. The most common reason for non-validated results was the result limit check (41%). A total of 328 reports evaluated by reviewers were compared to AV system. The statistical analysis resulted in a κ value between 0.39 and 0.63 (P < 0.001) and an agreement rate between 79% and 88%.
ConclusionsOur improved model can help laboratories design, build, and validate AV systems and be used as starting point for different test groups.
Publisher
Croatian Society for Medical Biochemistry and Laboratory Medicine
Subject
Biochemistry (medical),Clinical Biochemistry
Reference29 articles.
1. Sağlık Bakanlığı TC. Tıbbi laboratuvarda onay destek sistemi [Internet]. 2018 Mar [Cited 2021 Jul 9]. Available from: https://shgmtetkikdb.saglik.gov.tr/Eklenti/15045/0/tibbi-laboratuvarda-onay-desteek41916132pdf.pdf (in Turkish)
2. Establishing the reference change values (RCVs) and validating the delta check auto-verification in a clinical biochemistry laboratory.;Fernandez;Muller J Med Sci Res.,2017
3. Collection, evaluation, and transmission of hospital laboratory data.;Lindberg;Methods Inf Med,1967
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献