IntroductionBlood samples having inappropriate volume are a substantial part of preanalytical errors. Inadequate sample volume for glycated haemoglobin (HbA1c) test may be a common problem of patients with diabetes mellitus having vascular changes. In this study, we compared HbA1c concentrations of underfilled and appropriately filled blood collection tubes.
Materials and methodsTo compare HbA1c concentrations, blood samples were collected into 2 mL tubes containing K3-EDTA from 109 subjects. Two blood samples (underfilled and appropriately filled) were drawn from a patient by the same personnel and materials. HbA1c measurements were assayed on a Cobas 6000 analyser module c 501 (Roche Diagnostics, Mannheim, Germany). The HbA1c% results were compared by t-test and Wilcoxon’s signed-rank statistical methods (SPSS Inc., Chicago, USA). Bias analysis was performed using Microsoft Excel 4.0.
ResultsUnderfilled samples were classified three groups (group 1, N = 44; group 2, N = 36; and group 3, N = 29) according to the filling ratio of the samples; 0.5 mL and below (< 25%), 0.5-1.0 mL (25-50%), and 1.0-2.0 mL (> 50%), respectively. When we compared underfilled tubes with pairing filled tubes, there was a statistically significant difference only with tubes filled less than 25% (P = 0.030). Furthermore, we have done bias analysis between paired tubes according to the diagnostic cut-off value of 6.5%. The bias was more prominent in up to 50% underfilled blood tubes (1.1%), when HbA1c concentrations were below the diagnostic cut-off of 6.5%.
ConclusionsWe suggest that the blood tubes with EDTA for HbA1c measurement should be filled with at least 50% to avoid clinical variations.
Publisher
Croatian Society for Medical Biochemistry and Laboratory Medicine
Subject
Biochemistry (medical),Clinical Biochemistry
Reference14 articles.
1. Glycemic targets. Standards of medical care in diabetes-2015.;American Diabetes Association;Diabetes Care,2015
2. Blood sample quality.;Lippi;Diagnosis (Berl),2019
3. Preanalytical quality improvement: in quality we trust.;Lippi;Clin Chem Lab Med,2013
4. Phlebotomy issues and quality improvement in results of laboratory testing.;Lippi;Clin Lab,2006
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献