A Comparison of the BAX System Method to the U.S. Food and Drug Administration's Bacteriological Analytical Manual and International Organization for Standardization Reference Methods for the Detection of Salmonella in a Variety of Soy Ingredients

Author:

BELETE TAMRAT1,CROWLEY ERIN2,BIRD PATRICK2,GENSIC JOSEPH1,WALLACE F. MORGAN3

Affiliation:

1. 1DuPont Nutrition & Health (formerly Solae, LLC), 4300 Duncan Avenue, St. Louis, Missouri 63119

2. 2Q Laboratories, Inc., 1400 Harrison Avenue, Cincinnati, Ohio 45214

3. 3DuPont Nutrition & Health, ESL Building 400, Route 141 and Henry Clay Road, Wilmington, Delaware 19880, USA

Abstract

The performances of two DuPont BAX System PCR assays for detecting Salmonella on a variety of low-moisture soy ingredients were evaluated against the U.S. Food and Drug Administration's Bacteriological Analytical Manual (FDA BAM) method or the International Organization for Standardization (ISO) 6579 reference method. These evaluations were conducted as a single laboratory validation at an ISO 17025 accredited third-party laboratory. Validations were conducted on five soy ingredients: isolated soy protein (ISP), soy fiber, fluid soy lecithin, deoiled soy lecithin, and soy nuggets, using a paired-study design. The ISP was analyzed as both 25- and 375-g composite test portions, whereas all other sample matrices were analyzed as 375-g composite test portions. To evaluate 25-g test portions of ISP, the test material was inoculated using Salmonella enterica subsp. enterica serovar Mbandaka (Q Laboratories isolate 11031.1). Salmonella enterica subsp. enterica serovar Tennessee (Q Laboratories isolate 11031.3) was used for all other trials. For each trial of the method comparison, 25 samples were analyzed for each matrix: 5 uninoculated controls and 20 samples inoculated at low levels (0.2 to 2 CFU per test portion) that were targeted to achieve fractionally positive results (25 to 75%). Using McNemar's chi-square analysis, no significant difference at P ≥ 0.05 (χ2 ≤ 3.84) was observed between the number of positives obtained by the BAX System and the reference methods for all five test matrices evaluated. These studies indicate that the BAX System PCR assays, in combination with the single buffered peptone water primary enrichment and subsequent brain heart infusion regrowth step, demonstrate equivalent sensitivity and robustness compared with the FDA BAM and ISO reference methods for both 25- and 375-g composite samples. Moreover, there was no observed reduction of sensitivity in the larger 375-g composite samples for all five matrices.

Publisher

International Association for Food Protection

Subject

Microbiology,Food Science

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3