Use of ethylene-oxide gas sterilisation to terminate multidrug-resistant bacterial outbreaks linked to duodenoscopes

Author:

Muscarella Lawrence FORCID

Abstract

BackgroundCleaning and high-level disinfection have been the standard in the USA and other countries for reprocessing flexible endoscopes, including duodenoscopes and other types of gastrointestinal endoscopes. For decades, this practice has been a cornerstone for infection prevention in the endoscopic setting. However, amid recent reports associating the use of duodenoscopes with infections and outbreaks of carbapenem-resistant Enterobacteriaceae (CRE) and related multidrug-resistant organisms (MDROs), reasonable questions about the adequacy of current practices for reprocessing duodenoscopes have emerged.ObjectivesTo review and evaluate the adequacy of current reprocessing practices for preventing duodenoscopes from transmitting CRE and related MDROs.MethodsThe MEDLINE/PubMed database was searched to identify published cases associating confirmed (or suspected) infections of CRE or a related MDRO with exposure to a duodenoscope since 2012, when duodenoscopes became a recognised risk factor for the transmission of CRE. The Internet was also searched to identify news articles and other reports documenting eligible cases occurring during this same timeframe but not identified during the MEDLINE database’s search. The Food and Drug Administration’s (FDA) medical device database was queried to identify regulatory reports describing these same types of cases, also recorded since 2012. The clinical and reprocessing details of each eligible case were reviewed to identify (when possible): (a) the reprocessing method (e.g., high-level disinfection) performed at the time of the infections, (b) whether the facility’s compliance with the manufacturer’s reprocessing instructions was confirmed, and (c) the measure(s) or corrective action(s) the facility implemented to prevent additional multidrug-resistant infections.ResultsSeventeen cases in the USA and six in other countries (primarily Europe) associating infections (and colonizations) of CRE or a related MDRO with exposure to a duodenoscope were reviewed. Fourteen of these 23 outbreaks were caused by CRE, and six by a related MDRO. Two of these six latter cases identified Klebsiella pneumoniae carrying the mcr-1 gene as the pathogen. For 12 of these 23 cases, it was reported or implied that the duodenoscope was being high-level disinfected at the time of the infections, consistent with published guidelines. For the remaining 11 cases, the associated report(s) did not clearly identify how the duodenoscope was being reprocessed at the time of the infections (although it may be reasonably concluded that at least some, if not all, of these 11 cases involved high-level disinfection).Further, eight of the 23 cases reported the duodenoscope was being reprocessed in accordance with the manufacturer’s instructions for use (and professional guidelines) at the time of the infections. Seven of the cases discussed the design of the duodenoscope (eg, the forceps elevator mechanism) in the context of reprocessing and the infections. Three of the cases identified one or more reprocessing lapses, including inadequate cleaning, delayed reprocessing and improper drying and/or storage of the duodenoscope. Most of these 23 cases were associated with exposure to a duodenoscope model featuring a sealed elevator-wire channel. Six of the cases reported adopting (or in one case supplementing high-level disinfection with) ethylene oxide (EO) gas sterilisation of the duodenoscope, with at least three reporting this measure terminated the outbreak. Other measures adopted to prevent additional infections included removing the implicated duodenoscope from use, re-training staff about proper cleaning, microbiological culturing of the duodenoscope and returning the duodenoscope to the manufacturer for evaluation, maintenance and/or repair.ConclusionsThis study's findings suggest current reprocessing practices may not always be sufficiently effective to prevent a duodenoscope from transmitting CRE and related MDROs, at least in some circumstances including an outbreak setting. Factors this review identifed that may contribute to the device remaining contaminated after reprocessing include the device’s design; breaches of recommended reprocessing guidelines (eg, inadequate manual cleaning, delayed reprocessing or improper device storage); damage to the device; lacking servicing, maintenance or repair; and/or the presence of biofilms. Measures that can mitigate the impact of these and other reprocessing challenges and reduce, if not eliminate, the risk of transmission of CRE or a related MDRO by a duodenoscope include the use of EO gas sterilization (or another comparably effective process or method). In 2015, the FDA suggested healthcare facilities consider performing at least one of four supplemental measures, which include EO gas sterilisation, to improve the effectiveness of duodenoscope reprocessing. Whether the FDA and Centers for Disease Control and Prevention might reclassify duodenoscopes as critical devices requiring sterilisation is currently unresolved.

Funder

Andersen Products, Inc. Haw River, NC

Publisher

BMJ

Subject

Gastroenterology

Reference54 articles.

1. Centers for Disease Control and Prevention (CDC) . Guideline for disinfection and sterilization in healthcare facilities, 2008.

2. ASGE guideline for infection control during GI endoscopy;Calderwood;Gastrointest Endosc,2018

3. Society of Gastroenterology Nurses and Associates (SGNA) . Guideline for use of high-level disinfectants and sterilants for reprocessing flexible gastrointestinal endoscopes, 2013.

4. Spaulding . EH The role of chemical disinfection in the prevention of nosocomial infections. In: Brachman PS , Eickoff TC , eds. Proceedings of the International Conference on nosocomial infections, 1970. Chicago: American Hospital Association, 1971: 254–74.

5. Food and Drug Administration (FDA) . Reprocessing medical devices in health care settings: validation methods and labeling guidance for industry and food and drug administration staff, 2015: 1–40.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3