Evaluation of PET quantitation accuracy among multiple discovery IQ PET/CT systems via NEMA image quality test

Author:

Vallot DelphineORCID,De Ponti Elena,Morzenti Sabrina,Gramek Anna,Pieczonka Anna,Llompart Gabriel Reynés,Siennicki Jakub,Deak Paul,Dutta Chiranjib,Uribe Jorge,Caselles Olivier

Abstract

Abstract Introduction Quantitative imaging biomarkers are becoming usual in oncology for assessing therapy response. The harmonization of image quantitation reporting has become of utmost importance due to the multi-center trials increase. The NEMA image quality test is often considered for the evaluation of quantitation and is more accurate with a radioactive solid phantom that reduces variability. The goal of this project is to determine the level of variability among imaging centers if acquisition and imaging protocol parameters are left to the center’s preference while all other parameters are fixed including the scanner type. Methods A NEMA-IQ phantom filled with radioactive 68Ge solid resin was imaged in five clinical sites throughout Europe. Sites reconstructed data with OSEM and BSREM algorithms applying the sites’ clinical parameters. Images were analyzed according with the NEMA-NU2-2012 standard using the manufacturer-provided NEMA tools to calculate contrast recovery (CR) and background variability (BV) for each sphere and the lung error (LE) estimation. In addition, a 18F-filled NEMA-IQ phantom was also evaluated to obtain a gauge for variability among centers when the sites were provided with identical specific instructions for acquisition and reconstruction protocol (the aggregate of data from 12 additional sites is presented). Results The data using the 68Ge solid phantom showed no statistical differences among different sites, proving a very good reproducibility among the PET center models even if dispersion of data is higher with OSEM compared to BSREM. Furthermore, BSREM shows better CR and comparable BV, while LE is slightly reduced. Two centers exhibit significant differences in CR and BV values for the 18F NEMA NU2-2012 experiments; these outlier results are explained. Conclusion The same PET system type from the various sites produced similar quantitative results, despite allowing each site to choose their clinical protocols with no restriction on data acquisition and reconstruction parameters. BSREM leads to lower dispersion of quantitative data among different sites. A solid radioactive phantom may be recommended to qualify the sites to perform quantitative imaging.

Publisher

Springer Science and Business Media LLC

Subject

Radiology Nuclear Medicine and imaging,Instrumentation,Biomedical Engineering,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3