Investigation of image-based lesion and kidney dosimetry protocols for 177Lu-PSMA-I&T therapy with and without a late SPECT/CT acquisition

Author:

Resch Sandra,Takayama Fouladgar Sarah,Zacherl Mathias,Sheikh Gabriel T.,Liubchenko Grigory,Rumiantcev Mikhail,Unterrainer Lena M.,Wenter Vera,Bartenstein Peter,Ziegler Sibylle I.,Ilhan Harun,Beyer Leonie,Böning Guido,Delker Astrid

Abstract

Abstract Background 177Lu-PSMA therapy has been successfully used to prolong the survival of patients with metastatic castration-resistant prostate cancer. Patient-specific dosimetry based on serial quantitative SPECT/CT imaging can support the understanding of dose–effect relationships. However, multiple SPECT/CT measurements can be challenging for patients, which motivates the investigation of efficient sampling schedules and their impact on dosimetry. In this study, different time samplings with respect to the number and timing of SPECT/CT acquisitions with and without a late measurement were investigated. Materials and methods In total, 43 lesions and 10 kidneys of 5 patients receiving 177Lu-PSMA-I&T therapy were investigated. Whole-body SPECT/CT measurements were performed at 1, 2, 3 and 7 days post-injection. For both lesions (isocontour-based segmentation) and kidneys (CT-based segmentation), a reference model was employed including all four time points. To identify the best-matching fit function out of a pre-defined set of models, visual inspection, coefficients of variation and sum of squared errors were considered as goodness-of-fit criteria. Biologically effective doses (BEDs) calculated with different time samplings (days 1, 2, 3/1, 2, 7/1, 3, 7/2, 3, 7 and 1, 2/1, 3/1, 7) were compared to the reference. Results The best-fit function was found to be a mono-exponential model for lesions and a bi-exponential model with a population-based parameter and two free parameters for kidneys. The BEDs calculated with the time sampling 1, 3, 7 days showed the lowest deviations from the reference for lesions with 4 ± 5%. Without day 7, still 86% of all lesions showed deviations from the reference < 10%. The outlier deviations showed a positive correlation with the effective half-life of the respective lesions. For kidneys, including days 1, 2, 3 achieved the best results with 0 ± 1%. Generally, deviations for kidneys were found to be small for all time samplings (max. 13%). Conclusions For combined optimization of the SPECT/CT time sampling for kidney and lesion dosimetry during 177Lu-PSMA-I&T therapy, the sampling with days 1, 3, 7 showed the smallest deviation from the reference. Without a late acquisition, using the schedule with days 1, 2, 3 is likewise feasible.

Funder

Universitätsklinik München

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Biomedical Engineering,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3