Abstract
Abstract
Background
Quantitative whole-body PET/MRI relies on accurate patient-specific MRI-based attenuation correction (AC) of PET, which is a non-trivial challenge, especially for the anatomically complex head and neck region. We used a deep learning model developed for dose planning in radiation oncology to derive MRI-based attenuation maps of head and neck cancer patients and evaluated its performance on PET AC.
Methods
Eleven head and neck cancer patients, referred for radiotherapy, underwent CT followed by PET/MRI with acquisition of Dixon MRI. Both scans were performed in radiotherapy position. PET AC was performed with three different patient-specific attenuation maps derived from: (1) Dixon MRI using a deep learning network (PETDeep). (2) Dixon MRI using the vendor-provided atlas-based method (PETAtlas). (3) CT, serving as reference (PETCT). We analyzed the effect of the MRI-based AC methods on PET quantification by assessing the average voxelwise error within the entire body, and the error as a function of distance to bone/air. The error in mean uptake within anatomical regions of interest and the tumor was also assessed.
Results
The average (± standard deviation) PET voxel error was 0.0 ± 11.4% for PETDeep and −1.3 ± 21.8% for PETAtlas. The error in mean PET uptake in bone/air was much lower for PETDeep (−4%/12%) than for PETAtlas (−15%/84%) and PETDeep also demonstrated a more rapidly decreasing error with distance to bone/air affecting only the immediate surroundings (less than 1 cm). The regions with the largest error in mean uptake were those containing bone (mandible) and air (larynx) for both methods, and the error in tumor mean uptake was −0.6 ± 2.0% for PETDeep and −3.5 ± 4.6% for PETAtlas.
Conclusion
The deep learning network for deriving MRI-based attenuation maps of head and neck cancer patients demonstrated accurate AC and exceeded the performance of the vendor-provided atlas-based method both overall, on a lesion-level, and in vicinity of challenging regions such as bone and air.
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Instrumentation,Biomedical Engineering,Radiation
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献