Abstract
Abstract
Background
Halo artifacts from urinary catheters can occur due to inaccurate scatter correction, and the artifacts affect the tumor visibility in 18F-FDG PET/CT images. We investigated the incidence rate and the mechanisms of halo-artifact generation and explored several scatter correction techniques to prevent artifacts.
Methods
We conducted patient and phantom studies. (1) We retrospectively reviewed the cases of patients who had undergone 18F-FDG PET/CT scans. To determine the frequency of halo-artifact generation, we used the patients’ PET images with a standard scatter correction based on a tail-fitted single-scatter simulation (TF-SSS) using 4-mm voxel μ-maps (TFS 4-mm). (2) We performed phantom studies to evaluate the effects of a urine catheter and two scatter correction techniques, i.e., TF-SSS with 2-mm voxel μ-maps (TFS 2-mm) and a Monte Carlo-based single-scatter simulation (MC-SSS) using 4-mm voxel μ-maps (MCS 4-mm). The average standardized uptake values (SUVs) were measured for axial PET images. (3) Using the patients’ data, we investigated whether TFS 2-mm and MCS 4-mm can eliminate the artifacts in the clinical images.
Results
(1) There were 61 patients with urinary catheters; in five (8.2%), halo artifacts were observed in the TFS 4-mm PET images. (2) The phantom study clearly reproduced the halo artifacts in the TFS 4-mm PET images. The halo artifacts were generated when urine moved in the interval between the CT and PET imaging, and when the urinary catheter was placed in a circular shape. The SUVs for the TFS 4-mm and TFS-2mm PET images were underestimated at the halo-artifact regions, whereas the SUVs for the MCS 4-mm PET images were close to the true values. (3) The halo artifacts disappeared in the TFS 2-mm PET images in 4/5 patients but not 1/5 patient, whereas the halo artifacts were completely absent in the MCS 4-mm PET images in 5/5 patients.
Conclusions
These data suggest that halo artifacts are caused if the PET images do not correspond to the physical material in the μ-maps, which induces the scatter correction error. With the MC-SSS, it was possible to accurately estimate the scatter without generating halo artifacts.
Funder
The Ministry of Education,Culture,Sports,Science and Technology
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Instrumentation,Biomedical Engineering,Radiation
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献