Abstract
Abstract
Background
Dosimetry promises many advantages for radiopharmaceutical therapies but repeat post-therapy imaging for dosimetry can burden both patients and clinics. Recent applications of reduced time point imaging for time-integrated activity (TIA) determination for internal dosimetry following 177Lu-DOTATATE peptide receptor radionuclide therapy have shown promising results that allow for the simplification of patient-specific dosimetry. However, factors such as scheduling can lead to sub-optimal imaging time points, but the resulting impact on dosimetry accuracy is still under investigation. We use four-time point 177Lu SPECT/CT data for a cohort of patients treated at our clinic to perform a comprehensive analysis of the error and variability in time-integrated activity when reduced time point methods with various combinations of sampling points are employed.
Methods
The study includes 28 patients with gastroenteropancreatic neuroendocrine tumors who underwent post-therapy SPECT/CT imaging at approximately 4, 24, 96, and 168 h post-therapy (p.t.) following the first cycle of 177Lu-DOTATATE. The healthy liver, left/right kidney, spleen and up to 5 index tumors were delineated for each patient. Time-activity curves were fit with either monoexponential or biexponential functions for each structure, based on the Akaike information criterion. This fitting was performed using all 4 time points as a reference and various combinations of 2 and 3 time points to determine optimal imaging schedules and associated errors. 2 commonly used methods of single time point (STP) TIA estimation are also evaluated. A simulation study was also performed with data generated by sampling curve fit parameters from log-normal distributions derived from the clinical data and adding realistic measurement noise to sampled activities. For both clinical and simulation studies, error and variability in TIA estimates were estimated with various sampling schedules.
Results
The optimal post-therapy imaging time period for STP estimates of TIA was found to be 3–5 days (71–126 h) p.t. for tumor and organs, with one exception of 6–8 days (144–194 h) p.t. for spleen with one STP approach. At the optimal time point, STP estimates give mean percent errors (MPE) within ± 5% and SD < 9% across all structures with largest magnitude error for kidney TIA (MPE = − 4.1%) and highest variability also for kidney TIA (SD = 8.4%). The optimal sampling schedule for 2TP estimates of TIA is 1–2 days (21–52 h) p.t. followed by 3–5 days (71–126 h) p.t. for kidney, tumor, and spleen. Using the optimal sampling schedule, the largest magnitude MPE for 2TP estimates is 1.2% for spleen and highest variability is in tumor with SD = 5.8%. The optimal sampling schedule for 3TP estimates of TIA is 1–2 days (21–52 h) p.t. followed by 3–5 days (71–126 h) p.t. and 6–8 days (144–194 h) p.t. for all structures. Using the optimal sampling schedule, the largest magnitude MPE for 3TP estimates is 2.5% for spleen and highest variability is in tumor with SD = 2.1%. Simulated patient results corroborate these findings with similar optimal sampling schedules and errors. Many sub-optimal reduced time point sampling schedules also exhibit low error and variability.
Conclusions
We show that reduced time point methods can be used to achieve acceptable average TIA errors over a wide range of imaging time points and sampling schedules while maintaining low uncertainty. This information can improve the feasibility of dosimetry for 177Lu-DOTATATE and elucidate the uncertainty associated with non-ideal conditions.
Funder
National Cancer Institute
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging
Reference29 articles.
1. Strosberg J, El-Haddad G, Wolin E, et al. Phase 3 trial of 177Lu-dotatate for midgut neuroendocrine tumors. N Engl J Med. 2017;376(2):125–35.
2. Strosberg JR, Caplin ME, Kunz PL, et al. 177Lu-Dotatate plus long-acting octreotide versus high‑dose long-acting octreotide in patients with midgut neuroendocrine tumours (NETTER-1): final overall survival and long-term safety results from an open-label, randomised, controlled, phase 3 trial [published correction appears in Lancet Oncol. 2022 Feb;23(2):e59]. Lancet Oncol. 2021;22(12):1752–63.
3. Siegel JA, Thomas SR, Stubbs JB, et al. MIRD pamphlet no. 16: Techniques for quantitative radiopharmaceutical biodistribution data acquisition and analysis for use in human radiation dose estimates. J Nucl Med. 1999;40(2):37S-61S.
4. Hänscheid H, Lapa C, Buck AK, Lassmann M, Werner RA. Dose mapping after endoradiotherapy with 177Lu-DOTATATE/DOTATOC by a single measurement after 4 days. J Nucl Med. 2018;59(1):75–81.
5. Madsen MT, Menda Y, O’Dorisio TM, O’Dorisio MS. Technical Note: Single time point dose estimate for exponential clearance. Med Phys. 2018;45(5):2318–24.
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献