Abstract
Abstract
Objective
To present a case series of spontaneous structural failure of bridging stentgrafts (BSGs) after branched endovascular aortic repair (bEVAR), as well as their failure types and their detection. While bEVAR is a safe and effective procedure, one main limitation is the reintervention rate associated with the BSGs. Structural failure of BSGs, defined as fabric disruption, stent fracture with leak or complete separation is a major cause for reinterventions and difficult to detect in computed tomography angiography (CTA).
Methods
From a multicenter bEVAR complication database, structural BSG failures were identified. Patient and BSG characteristics, detection mode, failure type, treatment and outcome were recorded and compared with bEVAR patients with intact BSGs.
Results
Twenty-three BSG failures were detected in 12 patients with only 43% directly identified in CTA, after a mean of 21.4 months after implantation. The BSGs were Advanta (n = 4), E-Ventus (n = 16) and BeGraft (n = 3) in 10 renal, 9 superior mesenteric, and 4 celiac branches. Religning with another BSG was successful in 20/22 cases, one BSG failure necessitated renal branch embolization (organ loss), and one mesenteric bypass surgery. Two reintervention-related mortalities occurred.
Conclusion
Structural failure of BSGs is a serious limitation for bEVAR, which can result in high reintervention rates and serious complications. BSG failure typically occurs in single-layer types and events are clustered in patients. The necessary reinterventions carry serious morbidity and mortality. Since the use as BSG in bEVAR is off-label with all current BSG manufacturers, caution is advised regarding patient-informed consent.
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献