Reducing stiffness of shock-absorbing pylon amplifies prosthesis energy loss and redistributes joint mechanical work during walking

Author:

Maun Jenny Anne,Gard Steven A.,Major Matthew J.,Takahashi Kota Z.

Abstract

Abstract Background A shock-absorbing pylon (SAP) is a modular prosthetic component designed to attenuate impact forces, which unlike traditional pylons that are rigid, can compress to absorb, return, or dissipate energy. Previous studies found that walking with a SAP improved lower-limb prosthesis users’ comfort and residual limb pain. While longitudinal stiffness of a SAP has been shown to affect gait kinematics, kinetics, and work done by the entire lower limb, the energetic contributions from the prosthesis and the intact joints have not been examined. The purpose of this study was to determine the effects of SAP stiffness and walking speed on the mechanical work contributions of the prosthesis (i.e., all components distal to socket), knee, and hip in individuals with a transtibial amputation. Methods Twelve participants with unilateral transtibial amputation walked overground at their customary (1.22 ± 0.18 ms−1) and fast speeds (1.53 ± 0.29 ms−1) under four different levels of SAP stiffness. Power and mechanical work profiles of the leg joints and components distal to the socket were quantified. The effects of SAP stiffness and walking speed on positive and negative work were analyzed using two-factor (stiffness and speed) repeated-measure ANOVAs (α = 0.05). Results Faster walking significantly increased mechanical work from the SAP-integrated prosthesis (p < 0.001). Reducing SAP stiffness increased the magnitude of prosthesis negative work (energy absorption) during early stance (p = 0.045) by as much as 0.027 Jkg−1, without affecting the positive work (energy return) during late stance (p = 0.159), suggesting a damping effect. This energy loss was partially offset by an increase in residual hip positive work (as much as 0.012 Jkg−1) during late stance (p = 0.045). Reducing SAP stiffness also reduced the magnitude of negative work on the contralateral sound limb during early stance by 11–17% (p = 0.001). Conclusions Reducing SAP stiffness and faster walking amplified the prostheses damping effect, which redistributed the mechanical work, both in magnitude and timing, within the residual joints and sound limb. With its capacity to absorb and dissipate energy, future studies are warranted to determine whether SAPs can provide additional user benefit for locomotor tasks that require greater attenuation of impact forces (e.g., load carriage) or energy dissipation (e.g., downhill walking).

Funder

U.S. Department of Veterans Affairs, Rehabilitation Research and Development Service

Center of Research in Human Movement Variability, University of Nebraska at Omaha

Publisher

Springer Science and Business Media LLC

Subject

Health Informatics,Rehabilitation

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3