Abstract
AbstractMicroneedle electrode (ME) is used to monitor bioelectrical signals by penetrating via the skin, and it compensates for a limitation of surface electrodes. However, existing fabrication of ME have limited in controlling the shape of microneedles, which is directly relevant to the performance and durability of microneedles as an electrode. In this study, a novel method using 3D printing is developed to control needle bevel angles. By controlling the angle of printing direction, needle bevel angles are changed. Various angles of printing direction (0–90°) are investigated to fabricate moldings, and those moldings are used for microneedle fabrications using biocompatible polyimide (PI). The height implementation rate and aspect ratio are also investigated to optimize PI microneedles. The penetration test of the fabricated microneedles is conducted in porcine skin. The PI microneedle of 1000 μm fabricated by the printing angle of 40° showed the bevel angle of 54.5°, which can penetrate the porcine skin. The result demonstrates that this suggested fabrication can be applied using various polymeric materials to optimize microneedle shape.