Author:
Wu Zhongping,Huang Ruibo,Zhong Liping,Gao Yi,Zheng Jinping
Abstract
Abstract
Background
The spirometer is an important element in lung function examinations, and its accuracy is directly related to the accuracy of the results of these examinations and to the diagnosis and treatment of diseases. Our aim was to conduct a performance analysis of the detection techniques of differential pressure and ultrasonic portable spirometers commonly used in China.
Methods
A standard flow/volume simulator was used to analyze the performance (accuracy, repeatability, linearity, impedance, and so on) of portable spirometers, 4 imported and 6 domestic, based on 13 curves generated by different air sources in the ISO 26782:2009 standard. A Bland–Altman diagram was used to evaluate the consistency between the values measured by the spirometers and the simulator.
Results
The pass rates for accuracy, repeatability, linearity, and impedance for the 10 different portable spirometers were 50%, 100%, 70%, and 70%, respectively. Only 30% (3/10) of the spirometers—2 domestic and 1 imported—met all standards of quality and performance evaluation, while the rest were partially up to standard. In the consistency evaluation, only 3 spirometers were within both the consistency standard range and the acceptability range.
Conclusion
The quality and performance of different types of portable spirometers commonly used in the clinic differ. The use of a standard flow/volume simulator is helpful for the standard evaluation of the technical performance of spirometers.
Funder
National Key Technology R&D Program
National Science & Technology Pillar Program
Key Technology R&D Program of Zhejiang, China
Publisher
Springer Science and Business Media LLC
Subject
Pulmonary and Respiratory Medicine
Reference14 articles.
1. Jinping Z, Rongchang C. Pulmonary function-basic and clinical. Guangdong: Guangdong Science and Technology Press; 2007.
2. Pulmonary Function Professional Group, Respiratory Medicine Branch of Chinese Medical Association. Guidelines for pulmonary function examination (part 2)-spirometer examination. Chin J Tuberc Respir Med. 2014;37(7):481–486. https://doi.org/10.3760/cma.j.issn.1001-0939.2014.07.001.
3. National Research Center for Clinical Medicine of Respiratory Diseases. Lung function examination management norms. Beijing: People's Medical Publishing House, 2020.
4. ISO 26782 (2009), Anaesthetic and respiratory equipment —Spirometers intended for the measurement of time forced expired volumes in humans. https://www.iso.org/standard/43761.html. Published in July 2009.
5. Zhao Z, Wangjinping Q. Comparison of lung function measurement results with two different spirometers. Chin J Med Instrum. 2003:27(6):446–47. https://doi.org/10.3969/j.issn.1671-7104.2003.06.018.
Cited by
6 articles.
订阅此论文施引文献