Author:
Ding Jiarui,Shi Jinhong,Poirier Guy G,Wu Fang-Xiang
Abstract
Abstract
Background
Mass spectrometers can produce a large number of tandem mass spectra. They are unfortunately noise-contaminated. Noises can affect the quality of tandem mass spectra and thus increase the false positives and false negatives in the peptide identification. Therefore, it is appealing to develop an approach to denoising tandem mass spectra.
Results
We propose a novel approach to denoising tandem mass spectra. The proposed approach consists of two modules: spectral peak intensity adjustment and intensity local maximum extraction. In the spectral peak intensity adjustment module, we introduce five features to describe the quality of each peak. Based on these features, a score is calculated for each peak and is used to adjust its intensity. As a result, the intensity will be adjusted to a local maximum if a peak is a signal peak, and it will be decreased if the peak is a noisy one. The second module uses a morphological reconstruction filter to remove the peaks whose intensities are not the local maxima of the spectrum. Experiments have been conducted on two ion trap tandem mass spectral datasets: ISB and TOV. Experimental results show that our algorithm can remove about 69% of the peaks of a spectrum. At the same time, the number of spectra that can be identified by Mascot algorithm increases by 31.23% and 14.12% for the two tandem mass spectra datasets, respectively.
Conclusion
The proposed denoising algorithm can be integrated into current popular peptide identification algorithms such as Mascot to improve the reliability of assigning peptides to spectra.
Availability of the software
The software created from this work is available upon request.
Publisher
Springer Science and Business Media LLC
Subject
Molecular Biology,Biochemistry
Cited by
20 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献