Abstract
Abstract
Background
The bone-implant gap resulted from morphological mismatch between cervical bony endplates and implant footprint may have adverse impact on bone-implant interfacial osseointegration of cervical disc arthroplasty (CDA). The purpose of the study was to evaluate the impact of bone-implant gap size on the interfacial osseointegration in a rabbit animal model.
Methods
A series of round-plate implants with different teeth depth (0.5 mm, 1.0 mm, 1.5 mm and 2.0 mm) was specifically designed. A total of 48 New Zealand white rabbits were randomly categorized into four groups by the implants they received (0.5 mm: group A, 1.0 mm: group B, 1.5 mm: group C, 2.0 mm: group D). At 4th and 12th week after surgery, animals were sacrificed. Micro-CT, acid fuchsin and methylene blue staining and hematoxylin and eosin (HE) staining were conducted.
Results
At 4th week and 12th week after surgery, both micro-CT and HE staining showed more new bone formation and larger bone coverage in group A and group B than that in group C and group D. At 12th week, the bone biometric parameters were significantly superior in group C when compared with group D (p < 0.05). At 12th week, hard tissue slicing demonstrated larger portion of direct contact of new bone to the HA coating in group A and group B.
Conclusions
Bone-implant gap size larger than 1.0 mm negatively affected bone-implant osseointegration between compact bone and HA coated implant surface.
Funder
National Natural Science Foundation of China
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Rheumatology
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献