Mankinen M.,Virén T.,Seppälä J.,Koivumäki T.
Abstract
Abstract
Background
The dosimetric effect of setup uncertainty and tissue deformations in left-sided whole-breast irradiation with complementary surface-guided radiotherapy (SGRT) and cone-beam computed tomography (CBCT) setup was evaluated.
Method
Treatment courses of 40.05 Gy prescribed dose in 15 fractions were simulated for 29 patients by calculating the dose on deformed CT images, that were based on daily CBCT images, and deforming and accumulating the dose onto the planning CT image. Variability in clinical target volume (CTV) position and shape was assessed as the 95% Hausdorff distance (HD95) between the planning CTV and deformed CTV structures. DVH metrics were evaluated between the planned and simulated cumulative dose distributions using two treatment techniques: tangential volumetric modulated arc therapy (tVMAT) and conventional 3D-conformal radiotherapy (3D-CRT).
Results
Based on the HD95 values, the variations in CTV shape and position were enclosed by the 5 mm CTV-PTV margin in 85% of treatment fractions using complementary CBCT and SGRT setup. A residual error of 8.6 mm was observed between the initial SGRT setup and CBCT setup. The median CTV V95% coverage was 98.1% (range 93.1–99.8%) with tVMAT and 98.2% (range 84.5–99.7%) with 3D-CRT techniques with CBCT setup. With the initial SGRT-only setup, the corresponding coverages were 96.3% (range 92.6–99.4%) and 96.6% (range 84.2–99.4%), respectively. However, a considerable bias in vertical residual error between initial SGRT setup and CBCT setup was observed. Clinically relevant changes between the planned and cumulative doses to organs-at-risk (OARs) were not observed.
Conclusions
The CTV-to-PTV margin should not be reduced below 5 mm even with daily CBCT setup. Both tVMAT and 3D-CRT techniques were robust in terms of dose coverage to the target and OARs. Based on the shifts between setup methods, CBCT setup is recommended as a complementary method with SGRT.
Funder
Finnish Ministry of Social Affairs and Health
University of Jyväskylä
Publisher
Springer Science and Business Media LLC
Reference33 articles.
1. Radiotherapy Board. On Target 2: updated guidance for image-guided radiotherapy. London: The Royal College of Radiologists; 2021. Last accessed: 22nd June 2023. Available at: https://www.rcr.ac.uk/sites/default/files/radiotherapy-board-on-target-2-updated-guidance-image-guided-radiotherapy.pdf.
2. Luo J, Yin Z, Zhang Z, Guo X, Yu X, Wang J. Does the protocol-required uniform margin around the CTV adequately account for setup inaccuracies in whole breast irradiation? Radiat Oncol. 2021. https://doi.org/10.1186/s13014-021-01863-w.
3. Laaksomaa M, Sarudis S, Rossi M, Lehtonen T, Pehkonen J, Remes J, et al. AlignRT® and Catalyst™ in whole-breast radiotherapy with DIBH: Is IGRT still needed? J Appl Clin Med Phys. 2019. https://doi.org/10.1002/acm2.12553.
4. Jensen CA, Acosta Roa AM, Johansen M, Lund JÅ, Frengen J. Robustness of VMAT and 3DCRT plans toward setup errors in radiation therapy of locally advanced left-sided breast cancer with DIBH. Phys Med. 2018. https://doi.org/10.1016/j.ejmp.2017.11.019.
5. Wei X, Liu M, Ding Y, Li Q, Cheng C, Zong X, et al. Setup errors and effectiveness of Optical Laser 3D Surface imaging system (Sentinel) in postoperative radiotherapy of breast cancer. Sci Rep. 2018. https://doi.org/10.1038/s41598-018-25644-w.
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献