Abstract
Abstract
Purpose
To use cone-beam computed tomography (CBCT) imaging to determine the impacts of patient characteristics on the magnitude of geometric setup errors and obtain patient-specific planning target volume (PTV) margins from the correlated patient characteristics in whole breast irradiation (WBI).
Methods
Between January 2019 and December 2019, a total of 97 patients who underwent breast-conserving surgery, followed by intensity-modulated radiation therapy in WBI, were scanned with pre-treatment CBCT for the first three treatment fractions and weekly for the subsequent fractions. Setup errors in the left–right (LR), superior–inferior (SI) and anterior–posterior (AP) directions were recorded and analyzed with patient characteristics—including age, tumor location, body mass index (BMI), chest circumference (CC) and breast volume (BV)—to examine the predictors for setup errors and obtain specific PTV margins.
Results
A total of 679 CBCT images from 97 patients were acquired for analysis. The mean setup errors for the whole group were 2.32 ± 1.21 mm, 3.71 ± 2.21 mm and 2.75 ± 1.56 mm in the LR, SI and AP directions, respectively. Patients’ BMI, CC and BV were moderately associated with setup errors, especially in the SI directions (R = 0.40, 0.43 and 0.22, respectively). Setup errors in the SI directions for patients with BMI > 23.8 kg/m2, CC > 89 cm and BV > 657 cm3 were 4.56 ± 2.59 mm, 4.77 ± 2.42 mm and 4.30 ± 2.43 mm, respectively, which were significantly greater than those of patients with BMI ≤ 23.8 kg/m2, CC ≤ 89 cm and BV ≤ 657 cm3 (P < 0.05). Correspondingly, the calculated PTV margins in patients with BMI > 23.8 kg/m2, CC > 89 cm and BV > 657 cm3 were 4.25/7.95/4.93 mm, 4.37/7.66/5.24 mm and 4.22/7.54/5.29 mm in the LR/SI/AP directions, respectively, compared with 3.64/4.64/5.09 mm, 3.31/4.50/4.82 mm and 3.29/5.74/4.73 mm for BMI ≤ 23.8 kg/m2, CC ≤ 89 cm and BV ≤ 657 cm3, respectively.
Conclusions
The magnitude of geometric setup errors was moderately correlated with BMI, CC and BV. It was recommended to set patient-specific PTV margins according to patient characteristics in the absence of daily image-guided treatment setup.
Funder
National Natural Science Foundation of China
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Oncology
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献