Image-guided adaptive brachytherapy shows the ability to deliver high doses to tumors while sparing normal tissues. However, interfraction dose delivery introduces uncertainties to high dose estimation, which relates to normal tissue toxicity. The purpose of this study was to investigate the high-dose regions of two applicator approaches in brachytherapy.
Method
For 32 cervical cancer patients, the CT images from each fraction were wrapped to a reference image, and the displacement vector field (DVF) was calculated with a hybrid intensity-based deformable registration algorithm. The fractional dose was then accumulated to calculate the position and the overlap of high dose (D2cc) during multiple fractions.
Result
The overall Dice similarity coefficient (DSC) of the deformation algorithm for the bladder and the rectum was (0.97 and 0.91). No significant difference was observed between the two applicators. However, the location of the intracavitary brachytherapy (ICBT) high-dose region was relatively concentrated. The overlap volume of bladder and rectum D2cc was 0.42 and 0.71, respectively, which was higher than that of interstitial brachytherapy (ISBT) (0.26 and 0.31). The cumulative dose was overestimated in ISBT cases when using the GEC-recommended method. The ratio of bladder and rectum D2cc to the GEC method was 0.99 and 1, respectively, which was higher than that of the ISBT method (0.96 and 0.94).
Conclusion
High-dose regions for brachytherapy based on different applicator types were different. The 3D-printed ICBT has better high-dose region consistency than freehand ISBT and hence is more predictable.
Funder
natural science foundation of zhejiang province
zhejiang province key research tablrand development program
medical science and technology program of zhejiang province
National Natural Science Foundation of China
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Biomedical Engineering,General Medicine,Biomaterials,Radiological and Ultrasound Technology
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献