Relative position of the central hole after EVO-ICL implantation for moderate to high myopia

Author:

He Xiaojian,Niu Lingling,Miao Huamao,Zhao Feng,Zhou XingtaoORCID

Abstract

Abstract Background This study aimed to evaluate the relative position of the central hole (CH) of EVO Implantable Collamer Lens (EVO-ICL), the pupil center (CP), and the corneal center (CC) after implantation of EVO-ICLs for moderate to high myopia. Methods Eighty-nine eyes of forty-seven patients with moderate to high myopia were evaluated. The mean preoperative spherical equivalent (SE) was − 12.58 ± 4.13D. A routine postoperative follow-up was performed within 1 ~ 12 months. Positions of the CH of EVO-ICLs, the CP and the CC were recorded using a slit lamp anterior segment photography system, and their relative distances were calculated with the Visio image analysis software. Results All surgeries were performed safely, and no complications were observed in follow-ups 4.3 ± 4.82 months after surgery. At the last follow-up, the safety index (postoperative CDVA/preoperative CDVA) was 1.23 ± 0.48, and the efficacy index (postoperative UDVA/preoperative CDVA) was 1.08 ± 0.31. The CH in 85 eyes (95.51%) was superior to the CC, with 47.19% (42/89) on the temporal side and 48.31% (43/89) on the nasal side. The CH in 84 eyes (94.38%) was located on the temporal side of the CP, with 56.18% (50/89) superior and 38.2% (34/89) inferior to the CP. The CP of 85 eyes (95.51%) was superior on the nasal side of the CC. On the defined x-axis, the average distance from the CH to CC was significantly shorter than the average distance from the CP to CC (p < 0.001). Conclusions An imperfect match between the central hole of EVO-ICLs and the pupil center does not necessarily indicate ICL dislocation. Compared to the pupil center, the position of the central hole of EVO-ICL is closer to the corneal center.

Funder

National Natural Science Foundation of China

Project of Shanghai Science and Technology

Publisher

Springer Science and Business Media LLC

Subject

Ophthalmology,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3